IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 153521-001
Client Name, Sample Details

Coughy Bags
Taylor, MI 48180
Sample: Satelites
Type: Flower
Method: FE04U12 HPLC-UV

Test Conditions

Prepsheet ID#: MIP250211a
Scale: XS205-MI1
Temp: 22 °C
Baro Pressure: 980.3 hPa
Analyst: MEH
Technician: ARH

Sample ID#: 153521

Harvest/Process Date: 02/10/2025
Serving Size (g): 1
Date Received: 02/10/2025
Test Date: 02/11/2025
Valid Through: 02/12/2026
Report Issued: 02/13/2025

Sample #153521
Sample #153521 Sample #153521
Test
Compounds
Δ9-THC THCA Δ8-THC CBD CBDA CBG CBGA CBN CBC CBL THCV CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.3 21.7 N/D N/D N/D N/D 0.4 N/D N/D N/D N/D N/D 22.5 19.4 0.0 19.7
Amount (mg/g) 3.4 217.3 N/D N/D N/D N/D 4.01 N/D N/D N/D N/D N/D 224.71 194.0 0.0 197.5
Amount per Serving (mg) 3.4 217.3 N/D N/D N/D N/D 4.01 N/D N/D N/D N/D N/D 224.7 Serving Size~ (g): 1.0
LOQ (mg/g) 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74   %Decarb. THC CBD
±%RPD 0.93% 0.46% 0.01% 1.14% 1.49% 0.88% 1.36% 0.56% 0.37% 0.68% 1.70% 1.22% 2 N/A

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009.

FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products.

Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of Δ9-THC and Δ8-THC or CBD plus the amount of THC or CBD derived from THCA or CBDA, respectively.

These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value.

Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, Total CBG, CBC, THCV, CBDV, and CBL.

Total CBG is calculated as CBG plus CBGA*0.878 (the molar correction factor for CBGA to CBG conversion).

%Decarb. THC and CBD refer to the percentage of Δ9-THC or CBD relative to THCA or CBDA, respectively.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ARH - LM Iron laboratories ARH - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 1

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 153521-001
Client Name, Sample Details

Coughy Bags
Taylor, MI 48180
Sample: Satelites
Type: Flower
Method: FE04U12 HPLC-UV

Test Conditions

Prepsheet ID#: MIP250211a
Scale: XS205-MI1
Temp: 22 °C
Baro Pressure: 980.3 hPa
Analyst: MEH
Technician: ARH

Sample ID#: 153521

Harvest/Process Date: 02/10/2025
Serving Size (g): 1
Date Received: 02/10/2025
Test Date: 02/11/2025
Valid Through: 02/12/2026
Report Issued: 02/13/2025

Sample #153521
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Δ9-THC 0 5.16818 5.13193 99.30 5.0845 98.38 0.93%
THCA 0 4.93569 4.94695 100.23 4.96964 100.69 0.46%
Δ8-THC 0 5.28565 5.04534 95.45 5.0446 95.44 0.01%
CBD 0 5.15591 5.17956 100.46 5.12089 99.32 1.14%
CBDA 0 5.16081 5.13122 99.43 5.05533 97.96 1.49%
CBG 0 5.02664 4.98504 99.17 4.94142 98.30 0.88%
CBGA 0 5.14464 5.07541 98.65 5.00691 97.32 1.36%
CBN 0 5.21769 5.1122 97.98 5.08363 97.43 0.56%
CBC 0 5.03877 5.02532 99.73 5.00694 99.37 0.37%
CBCA N/A 0 0 0.00 0 0.00 0% N/A
CBL 0 5.33085 5.04109 94.56 5.00681 93.92 0.68%
THCV 0 4.98477 5.00982 100.50 4.92534 98.81 1.70%
CBDV 0 7.48636 5.05483 67.52 4.99341 66.70 1.22% LR

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ARH - LM Iron laboratories ARH - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 1