IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089
***Water Activity: 0.83

Test Conditions

Prepsheet ID#: MIP230717A
Scale: XS205-MI2
Temp: 21.5 °C
Baro Pressure: 978.6 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/17/2024
Report Issued: 07/18/2023

Sample #141413
Sample #141413 Sample #141413
Test
Compounds
Δ9-THC THCA Δ8-THC CBD CBDA CBG CBGA CBN CBC CBL THCV CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.3 N/D N/D N/D N/D N/D N/D N/D N/D N/D N/D N/D 0.3 0.3 0.0 0.3
Amount (mg/g) 2.7 N/D N/D N/D N/D 0.1 N/D N/D N/D N/D N/D N/D 2.8 2.7 0.0 2.8
Amount per Serving (mg) 10.233 N/D N/D N/D N/D 0.379 N/D N/D N/D N/D N/D N/D 10.6 Serving Size~ (g): 3.8
LOQ (mg/g) 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04 0.04   %Decarb. THC CBD
±%RPD 0.96% 1.17% 1.34% 0.52% 0.29% 0.44% 0.55% 0.96% 0.09% 1.89% 0.42% 0.02% 100% N/A

Serving size = 4.13 mL = 3.79g. 20 servings per container.. Sample passes for water activity (sample fails if Aw>0.85).

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009.

FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products.

Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of Δ9-THC and Δ8-THC or CBD plus the amount of THC or CBD derived from THCA or CBDA, respectively.

These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value.

Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, Total CBG, CBC, THCV, CBDV, and CBL.

Total CBG is calculated as CBG plus CBGA*0.878 (the molar correction factor for CBGA to CBG conversion).

%Decarb. THC and CBD refer to the percentage of Δ9-THC or CBD relative to THCA or CBDA, respectively.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089
***Water Activity: 0.83

Test Conditions

Prepsheet ID#: MIP230717A
Scale: XS205-MI2
Temp: 21.5 °C
Baro Pressure: 978.6 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/17/2024
Report Issued: 07/18/2023

Sample #141413
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Δ9-THC 0 5.25726 5.26353 100.12 5.21323 99.16 0.96%
THCA 0 5.118 5.05424 98.75 5.11369 99.92 1.17%
Δ8-THC 0 5.29002 5.36314 101.38 5.292 100.04 1.34%
CBD 0 5.37071 5.33904 99.41 5.31159 98.90 0.52%
CBDA 0 5.14804 5.18124 100.64 5.16624 100.35 0.29%
CBG 0 5.11766 5.16878 101.00 5.19181 101.45 0.44%
CBGA 0 4.978 5.02262 100.90 4.9951 100.34 0.55%
CBN 0 5.25964 5.28752 100.53 5.23723 99.57 0.96%
CBC 0 5.17078 5.17384 100.06 5.17854 100.15 0.09%
CBCA N/A 0 0 0.00 0 0.00 0%
CBL 0 5.01891 5.01313 99.88 4.91949 98.02 1.89%
THCV 0 5.29272 5.25682 99.32 5.27902 99.74 0.42%
CBDV 0 5.29938 5.24806 99.03 5.24716 99.01 0.02%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Pesticide Analysis by
Mass Spectrometer

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIPS230717a
Scale: XS205-MI1
Temp: 22.4 °C
Baro PE: 975.0 hPa
Analyst: MEH
Technician: MK

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/19/2023

Sample #141413
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
Aldicarb 0.400 0.125 Pass/<LOD Abamectin**** 0.500 0.125 Pass/<LOD
Acephate 0.400 0.125 Pass/<LOD Acequinocyl 2.000 0.250 Pass/<LOD
Acetamiprid 0.200 0.125 Pass/<LOD Azoxystrobin 0.200 0.125 Pass/<LOD
Bifenazate 0.200 0.125 Pass/<LOD Bifenthrin 0.200 0.125 Pass/<LOD
Boscalid 0.400 0.125 Pass/<LOD Carbaryl 0.200 0.125 Pass/<LOD
Carbofuran 0.200 0.125 Pass/<LOD Chlorantraniliprole 0.200 0.125 Pass/<LOD
Chlorfenapyr 1.000 0.500 Pass/<LOD Chlorpyrifos 0.200 0.125 Pass/<LOD
Clofentezine 0.200 0.125 Pass/<LOD Cyfluthrin** 1.000 0.500 Pass/<LOD
Cypermethrin*** 1.000 0.500 Pass/<LOD Daminozide 1.000 0.500 Pass/<LOD
DDVP (Dichlorvos) 1.000 0.250 Pass/<LOD Diazinon 0.200 0.125 Pass/<LOD
Dimethoate 0.200 0.125 Pass/<LOD Ethoprophos 0.200 0.125 Pass/<LOD
Etofenprox 0.400 0.125 Pass/<LOD Etoxazole 0.200 0.125 Pass/<LOD
Fenoxycarb 0.200 0.125 Pass/<LOD Fenpyroximate 0.400 0.125 Pass/<LOD
Fipronil 0.400 0.125 Pass/<LOD Flonicamid 1.000 0.125 Pass/<LOD
Fludioxonil 0.400 0.125 Pass/<LOD Hexythiazox 1.000 0.125 Pass/<LOD
Imazalil 0.200 0.125 Pass/<LOD Imidacloprid 0.400 0.125 Pass/<LOD
Kresoxim Methyl 0.400 0.125 Pass/<LOD Malathion 0.200 0.125 Pass/<LOD
Metalaxyl 0.200 0.125 Pass/<LOD Methiocarb 0.200 0.125 Pass/<LOD
Methomyl 0.400 0.125 Pass/<LOD Methyl Parathion 0.200 0.125 Pass/<LOD
MGK-264‡ 0.200 0.125 Pass/<LOD Myclobutanil 0.200 0.125 Pass/<LOD
Naled 0.500 0.125 Pass/<LOD Oxamyl 1.000 0.125 Pass/<LOD
Paclobutrazol 0.400 0.125 Pass/<LOD Permethrins† 0.200 0.125 Pass/<LOD
Phosmet 0.200 0.125 Pass/<LOD Piperonyl Butoxide 100.000 1.900 Not Tested
Prallethrin 0.200 0.125 Pass/<LOD Propiconazole 0.400 0.125 Pass/<LOD
Propoxur 0.200 0.125 Pass/<LOD Pyrethrins* 1.000 0.125 Pass/<LOD
Pyridaben 0.200 0.125 Pass/<LOD Spinosad***** 0.200 0.125 Pass/<LOD
Spiromesifen 0.200 0.125 Pass/<LOD Spirotetramat 0.200 0.125 Pass/<LOD
Spiroxamine‡ 0.400 0.125 Pass/<LOD Tebuconazole 0.400 0.125 Pass/<LOD
Thiacloprid 0.200 0.125 Pass/<LOD Thiamethoxam 0.200 0.125 Pass/<LOD
Trifloxystrobin 0.200 0.125 Pass/<LOD

* Pyrethrins are reported as the sum of Jasmolin I, Cinerin I, and Pyrethrin I
** Cyfluthrins are reported as the sum of isomers Cyfluthrin I, II, III, and IV
*** Cypermethrins are reported as the sum of isomers Cypermethrin I, II, III, and IV
**** Abamectin is reported as the sum of Avermectin B1a and Avermectin B1b
***** Spinosad is reported as the sum of Spinosyn A and Spinosyn D
† Permethrin and Prallethrin are reported as the sum of cis and trans isomers
‡ MGK-264 and Spiroximine are reported as the sum of isomers I and II
MRL - Maximum Residue Limit; LOD - Limit of Detection

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Pesticide QC Report

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIPS230717a
Scale: XS205-MI1
Temp: 22.4 °C
Baro PE: 975.0 hPa
Analyst: MEH
Technician: MK

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/19/2023

Sample #141413
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS recovery% MSD recovery% Relative Percent Difference (%) QC Flag
Acephate N.D. 1 0.803 0.795 80.30 79.50 1.00
Acequinocyl N.D. 1 0.713 0.703 71.30 70.30 1.41
Acetamiprid N.D. 1 0.863 0.811 86.30 81.10 6.21
Aldicarb N.D. 1 0.885 0.904 88.50 90.40 2.12
Avermectin B1a N.D. 1 0.716 0.713 71.60 71.30 0.42
Azoxystrobin N.D. 1 0.706 0.689 70.60 68.90 2.44
Bifenazate N.D. 1 0.783 0.751 78.30 75.10 4.17
Bifenthrin N.D. 1 0.814 0.751 81.40 75.10 8.05
Boscalid N.D. 1 0.763 0.759 76.30 75.90 0.53
Carbaryl N.D. 1 0.919 0.905 91.90 90.50 1.54
Carbofuran N.D. 1 0.891 0.86 89.10 86.00 3.54
Chlorantraniliprole N.D. 1 0.542 0.473 54.20 47.30 13.60 LR
Chlorfenapyr N.D. 1 0.837 0.796 83.70 79.60 5.02
Chlorpyriphos N.D. 1 0.722 0.711 72.20 71.10 1.54
Clofentezine N.D. 1 0.72 0.735 72.00 73.50 2.06
Cyfluthrin N.D. 1 0.813 0.787 81.30 78.70 3.25
Cypermethrin N.D. 1 0.787 0.712 78.70 71.20 10.01
Daminoside N.D. 1 0.651 0.626 65.10 62.60 3.92
Diazanon N.D. 1 0.992 0.979 99.20 97.90 1.32
Dichlorvos N.D. 1 0.769 0.721 76.90 72.10 6.44
Dimethoate N.D. 1 0.786 0.755 78.60 75.50 4.02
Ethoprophos N.D. 1 0.815 0.794 81.50 79.40 2.61
Etofenprox N.D. 1 0.797 0.76 79.70 76.00 4.75
Etoxazole N.D. 1 0.912 0.866 91.20 86.60 5.17
Fenoxycarb N.D. 1 0.764 0.727 76.40 72.70 4.96
Fenpyroximate N.D. 1 0.827 0.797 82.70 79.70 3.69
Fipronil N.D. 1 1.14 1.18 114.00 118.00 3.45
Flonicamid N.D. 1 0.763 0.718 76.30 71.80 6.08
Fludioxonil N.D. 1 1.24 1.29 124.00 129.00 3.95 HR
Hexythiazox N.D. 1 0.745 0.729 74.50 72.90 2.17
Imazalil N.D. 1 0.701 0.746 70.10 74.60 6.22
Imidacloprid N.D. 1 0.848 0.798 84.80 79.80 6.08
Kresoxim-methyl N.D. 1 0.896 0.913 89.60 91.30 1.88
Malathion N.D. 1 0.786 0.773 78.60 77.30 1.67
Metalaxyl N.D. 1 0.866 0.81 86.60 81.00 6.68
Methiocarb N.D. 1 0.859 0.822 85.90 82.20 4.40
Methomyl N.D. 1 0.906 0.857 90.60 85.70 5.56
MGK-264 N.D. 1 0.776 0.754 77.60 75.40 2.88
Myclobutanil N.D. 1 0.81 0.742 81.00 74.20 8.76
Naled (dibrom) N.D. 1 0.417 0.415 41.70 41.50 0.48 LR
Oxamyl N.D. 1 0.819 0.788 81.90 78.80 3.86
Paclobutrazol N.D. 1 0.835 0.819 83.50 81.90 1.93
Parathion-methyl N.D. 1 1.18 1.17 118.00 117.00 0.85
Permethrins N.D. 1 0.774 0.729 77.40 72.90 5.99
Phosmet N.D. 1 0.829 0.795 82.90 79.50 4.19
Piperonyl butoxide N/A N/A N/A N/A nan nan N/A
Prallethrin N.D. 1 0.768 0.753 76.80 75.30 1.97
Propiconazole N.D. 1 0.768 0.763 76.80 76.30 0.65
Propoxur N.D. 1 0.862 0.84 86.20 84.00 2.59
Pyrethrin N.D. 1 0.754 0.723 75.40 72.30 4.20
Pyridaben N.D. 1 0.83 0.798 83.00 79.80 3.93
SpinosynA N.D. 0.82 0.545 0.524 66.46 63.90 3.93
SpinosynD N.D. 0.18 0.17 0.154 94.44 85.56 9.88
Spiromesifen N.D. 1 0.338 0.326 33.80 32.60 3.61 LR
Spirotetramat N.D. 1 0.73 0.722 73.00 72.20 1.10
Spiroxamine N.D. 1 0.799 0.757 79.90 75.70 5.40
Tebuconazole N.D. 1 0.878 0.83 87.80 83.00 5.62
Thiacloprid N.D. 1 0.896 0.852 89.60 85.20 5.03
Thiamethoxam N.D. 1 0.796 0.765 79.60 76.50 3.97
Trifloxystrobin N.D. 1 0.788 0.75 78.80 75.00 4.94

N.D. = Not Detected

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 30%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (60-120%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 3 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Residual Solvents Analysis by
Mass Spectrometer

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: SOP FE-44-MI
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIHS230717A
Scale: XS205-MI1
Temp: 22.9 °C
Baro PE: 976.9 hPa
Analyst: MEH
Technician: MEH

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/19/2023

Sample #141413
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
1,2-Dichloroethane 5 1 Pass/<LOD 1,2-Dimethoxyethane 5,000 250 Not Tested
1,4-dioxane 5,000 25 Not Tested 1-Butanol 5,000 250 Not Tested
1-Pentanol 5,000 250 Not Tested 1-Propanol 5,000 250 Not Tested
2,2-Dimethylpropane (Neopentane) 5,000 25 Pass/<LOD 2,2-Dimethylbutane (Hexanes) 290 25 Pass/<LOD
2,3-Dimethylbutane (Hexanes) 290 25 Pass/<LOD 2-Butanol 5,000 25 Not Tested
2-Butanone (MEK) 5,000 250 Not Tested 2-Ethoxyethanol 5,000 25 Not Tested
2-Methylbutane (Isopentane) 5,000 25 Pass/<LOD 2-Methylpentane (Hexanes) 290 25 Pass/<LOD
2-Methylpropane (Isobutane) 5,000 25 Pass/<LOD 2-propanol (Isopropyl Alcohol) 5,000 25 Pass/<LOD
2-Propanone (Acetone) 5,000 25 Pass/<LOD 3-Methylpentane (Hexanes) 290 25 Pass/<LOD
Acetonitrile 410 30 Pass/<LOD Benzene 2 1 Pass/<LOD
Butane 5,000 25 Pass/<LOD Chloroform 60 1 Pass/<LOD
Cumene 5,000 25 Not Tested Cyclohexane 290 25 Not Tested
Dichloromethane 600 25 Pass/<LOD Dimethylsulfoxide (DMSO) 5,000 250 Not Tested
Ethanol 5,000 50 Pass/<LOD Ethyl acetate 5,000 25 Pass/<LOD
Ethyl ether 5,000 25 Pass/<LOD Ethylene glycol 5,000 25 Not Tested
Ethylene oxide 50 3 Pass/<LOD Heptane 5,000 25 Pass/<LOD
Hexane 290 25 Pass/<LOD Isopropyl acetate 5,000 25 Not Tested
Methanol 3,000 125 Pass/<LOD Naptha 400 100 Not Tested
N,N-Dimethylacetamide 5,000 250 Not Tested N,N-Dimethylformamide (DMF) 5,000 250 Not Tested
Pentane 5,000 25 Pass/<LOD Petroleum Ether 400 100 Not Tested
Propane 5,000 25 Pass/<LOD Pyridine 5,000 250 Not Tested
Sulfolane 5,000 25 Not Tested Tetrahydrofuran (THF) 5,000 25 Not Tested
Toluene 890 25 Pass/<LOD Trichloroethylene 80 25 Pass/<LOD
Xylenes* 2,170 25 Pass/<LOD

Total Butanes = 0 ug/g (PASS); Total Pentanes = 0 ug/g (PASS); Total Hexanes = 0 ug/g (PASS)

* Xylenes are reported as the sum of o-xylene, m-xylene, p-xylene, and ethylbenzene
MRL - Maximum Residue Limit; LOD - Limit of Detection

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 4 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Residual Solvents QC Report

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: SOP FE-44-MI
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIHS230717A
Scale: XS205-MI1
Temp: 22.9 °C
Baro PE: 976.9 hPa
Analyst: MEH
Technician: MEH

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/19/2023

Sample #141413
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS Recovery (%) MSD Recovery (%) Relative Percent Difference (%) QC Flag
1,2-Dichloroethane N.D. 250 293 290 117.20 116.00 1.03
n-Propane N.D. 250 176 188 70.40 75.20 6.59
Isobutane N.D. 250 178 177 71.20 70.80 0.56
Methanol N.D. 250 278 285 111.20 114.00 2.49
n-Butane N.D. 250 210 212 84.00 84.80 0.95
2,2-Dimethylpropane N.D. 250 211 212 84.40 84.80 0.47
Ethanol N.D. 250 277 274 110.80 109.60 1.09
Isopentane N.D. 250 248 246 99.20 98.40 0.81
Acetonitrile N.D. 250 285 285 114.00 114.00 0.00
Diethyl ether N.D. 250 282 279 112.80 111.60 1.07
2-propanone N.D. 250 293 295 117.20 118.00 0.68
2-propanol N.D. 250 280 284 112.00 113.60 1.42
n-Pentane N.D. 250 232 250 92.80 100.00 7.47
2,2-Dimethylbutane N.D. 250 279 266 111.60 106.40 4.77
2,3-Dimethylbutane N.D. 250 270 275 108.00 110.00 1.83
Methylene chloride N.D. 250 285 286 114.00 114.40 0.35
2-Methylpentane N.D. 250 270 275 108.00 110.00 1.83
3-Methylpentane N.D. 250 275 274 110.00 109.60 0.36
2-Butanol N.D. 0 0 0 0.00 0.00 0.00
n-Hexane N.D. 250 272 275 108.80 110.00 1.10
Ethyl acetate N.D. 250 286 288 114.40 115.20 0.70
Tetrahydrofuran N.D. 0 0 0 0.00 0.00 0.00
Trichloroethylene N.D. 250 293 294 117.20 117.60 0.34
Isopropyl acetate N.D. 0 0 0 0.00 0.00 0.00
Benzene N.D. 250 296 294 118.40 117.60 0.68
Chloroform N.D. 250 294 296 117.60 118.40 0.68
Cyclohexane N.D. 0 0 0 0.00 0.00 0.00
Ethylene glycol N.D. 0 0 0 0.00 0.00 0.00
Heptane N.D. 250 280 284 112.00 113.60 1.42
2-Ethoxyethanol N.D. 0 0 0 0.00 0.00 0.00
1,4-Dioxane N.D. 0 0 0 0.00 0.00 0.00
Pyridine N.D. 0 0 0 0.00 0.00 0.00
Toluene N.D. 250 298 301 119.20 120.40 1.00
Chlorobenzene N.D. 0 0 0 0.00 0.00 0.00
Ethylbenzene N.D. 0 0 0 0.00 0.00 0.00
m-Xylene/p-Xylene N.D. 500 595 606 119.00 121.20 1.83
o-Xylene N.D. 250 298 304 119.20 121.60 1.99
Cumene N.D. 0 0 0 0.00 0.00 0.00
Ethylene oxide N.D. 250 285 292 114.00 116.80 2.43

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

N.D. = Not Detected

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 5 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Microbiology Analysis

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Scale: XS205-MI1
Temp: 22.1 °C
Baro Pressure: 965.7 hPa
Analyst: SKM
Technician: SKM

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/20/2023

Sample #141413
Test Method MCFU (CFU/g) LOD (CFU/g) Status (CFU/g)
Total Aerobic Not Tested N/A 0 Not Tested
Yeast & Mold FE-66 (MPN) 10,001 100 Pass/<LOD
Bile-tolerant gram-negative bacteria Not Tested N/A 0 Not Tested
Total Coliforms FE-66 (MPN) 101 100 Pass/<LOD
E. Coli FE-65 (PCR) 1 1 Pass/Not Present
Salmonella FE-65 (PCR) 1 1 Pass/Not Present
Aspergillus spp. FE-65 (PCR) 1 1 Pass/Not Present

E. coli test screens exclusively for Shiga toxin-producing E. coli. TNTC values for yeast and mold are >150,000 CFU/g. TNTC values for total coliforms are >15,000 CFU/g. The maximum relative uncertainty is 6% for PCR and 40% for MPN.

TNTC = Too Numerous To Count. Maximum Colony Forming Units (MCFU) represents the minimum value that a test would need to achieve in order to be considered a failed result (if applicable). LOD is the Limit of Detection for the method. CFU/g = Colony Forming Units per gram. Aspergillus spp. consists of A. flavus, A. fumigatus, A. niger, and A. terreus.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 6 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Bio Analysis by
Microscopic Visual Inspection

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE02 Microscope Dissection
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Analyst: ARH
Technician: ARH

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/18/2023

Sample #141413
Sample #141413 Sample #141413
Severity Rating
Mold and Mildew 0 Pass
Mites (not Russet) 0 Pass
Russet Mites 0 Pass
Thrips 0 Pass
Aphids 0 Pass
Other Pests 0 Pass
Other Organic Matter 0 Pass
Inorganic Matter 0 Pass
Total Organic Foreign Matter <2% Pass
Inorganic Foreign Matter 0% Pass
Stem Material <5% Pass
Severity Rating Key
Correlating %FM ≤2% FM >2% FM
Severity Rating 0 None 1 Mild 2 Moderate 3 Severe Units
Mold and Mildew <2 2 3 4+ Locations/g
Mites (not russet) <1 1-5 6-10 11+ Amount/g
Russet Mites <1 1-10 11-50 51+ Amount/g
Thrips <1 1-3 4-5 6+ Amount/g
Aphids <1 1 2-4 5+ Amount/g
Other Pests <1 1-5 6-10 11+ Amount/g
Other Organic Matter <1 1-5 6-10 11+ Amount/g
Inorganic Matter 0 >0 Amount/g

In compliance with the MMFLA regulation, the action limits for foreign matter are as follows; no more that 2% organic matter, 5% stem material, and no presence of inorganic matter. Percent foreign contamination is based off of total surface area.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 7 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Heavy Metals Analysis

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE60 ICPMS-MI
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIICP 230717gMK
Temp: 22.4 °C
Baro PE: 975.0 hPa
Analyst: MK
Technician: MK

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/17/2023

Sample #141413
Sample #141413 Sample #141413
Test Compound MRL (PPM) LOD (PPM) Status (PPM)
Arsenic 1.5 0.1 Pass/<LOD
Copper 0 1.5 Not Tested
Cadmium 0.5 0.1 Pass/<LOD
Chromium 2 0.3 Pass/<LOD
Lead 0.5 0.25 Pass/<LOD
Mercury 3 0.05 Pass/<LOD
Nickel 0 0.25 Not Tested

MRL - Maximum Residue Limit; LOD - Limit of Detection

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 8 of 8

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 141413-001
Client Name, Sample Details

Northern Connections / Truu Cannabis (AU-P-000247)
Lansing , MI 48910
Sample: Green Apple Liquid Loud
License: AU-P-000247
Type: Infused Product
Method: FE60 ICPMS-MI
Metrc Test Pkg#: 1A4050300029041000005182
Metrc Source Pkg#: 1A4050300029041000005089

Test Conditions

Prepsheet ID#: MIICP 230717gMK
Temp: 22.4 °C
Baro PE: 975.0 hPa
Analyst: MK
Technician: MK

Sample ID#: 141413

Harvest/Process Date: 07/14/2023
Serving Size (g): 3.79
Date Received: 07/14/2023
Test Date: 07/17/2023
Valid Through: 07/16/2024
Report Issued: 07/17/2023

Sample #141413
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD QC Flag
Arsenic N/D 0.160 0.176 110.00 0.172 107.50
Cadmium N/D 0.160 0.176 110.00 0.173 108.13
Chromium N/D 0.480 0.535 111.46 0.529 110.21
Copper N/D 2.400 2.535 105.63 2.414 100.58
Lead N/D 0.400 0.426 106.50 0.424 106.00
Mercury N/D 0.080 0.093 116.25 0.095 118.75
Nickel N/D 0.400 0.514 128.50 0.512 128.00

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 9 of 8