IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082
***Water Activity: 0.43

Test Conditions

Prepsheet ID#: MIP221220B
Scale: XS205-MI2
Temp: 23 °C
Baro Pressure: 990.7 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/21/2023
Report Issued: 12/21/2022

Sample #139036
Sample #139036 Sample #139036
Test
Compounds
Δ9-THC THCA Δ8-THC CBD CBDA CBG CBGA CBN CBC CBL THCV CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.5 24.4 N/D N/D N/D 0.1 0.9 N/D N/D 0.1 N/D N/D 26.0 21.9 0.0 22.9
Amount (mg/g) 5.0 244.1 N/D N/D N/D 1.2 8.62 N/D N/D 0.8 N/D N/D 259.72 219.1 0.0 228.6
Amount per Serving (mg) 5 244.1 N/D N/D N/D 1.2 8.62 N/D N/D 0.8 N/D N/D 259.7 Serving Size~ (g): 1.0
LOQ (mg/g) 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74 0.74   %Decarb. THC CBD
±%RPD 1.63% 2.21% 1.17% 0.01% 0.15% 0.24% 0.02% 0.13% 1.61% 9.27% 0.07% 0.26% 2 N/A

Sample passes for water activity (sample fails if Aw>0.65).

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009.

FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products.

Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of Δ9-THC and Δ8-THC or CBD plus the amount of THC or CBD derived from THCA or CBDA, respectively.

These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value.

Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, Total CBG, CBC, THCV, CBDV, and CBL.

Total CBG is calculated as CBG plus CBGA*0.878 (the molar correction factor for CBGA to CBG conversion).

%Decarb. THC and CBD refer to the percentage of Δ9-THC or CBD relative to THCA or CBDA, respectively.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082
***Water Activity: 0.43

Test Conditions

Prepsheet ID#: MIP221220B
Scale: XS205-MI2
Temp: 23 °C
Baro Pressure: 990.7 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/21/2023
Report Issued: 12/21/2022

Sample #139036
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Δ9-THC 0 5.33682 5.31611 99.61 5.23022 98.00 1.63%
THCA 0 4.86884 5.06406 104.01 4.95314 101.73 2.21%
Δ8-THC 0 5.66089 5.52918 97.67 5.46489 96.54 1.17%
CBD 0 4.78718 4.94663 103.33 4.94597 103.32 0.01%
CBDA 0 4.76105 4.86214 102.12 4.8549 101.97 0.15%
CBG 0 5.32514 5.42985 101.97 5.44311 102.22 0.24%
CBGA 0 5.29509 5.40967 102.16 5.40875 102.15 0.02%
CBN 0 4.89847 4.85717 99.16 4.86372 99.29 0.13%
CBC 0 5.17193 5.30828 102.64 5.22368 101.00 1.61%
CBCA N/A 0 0 0.00 0 0.00 0%
CBL 0 5.559 5.69248 102.40 5.18792 93.32 9.27%
THCV 0 5.05315 5.22943 103.49 5.23307 103.56 0.07%
CBDV 0 5.18166 5.34231 103.10 5.35642 103.37 0.26%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Pesticide Analysis by
Mass Spectrometer

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Prepsheet ID#: MIPS221220a
Scale: XS205-MI1
Temp: 22.4 °C
Baro PE: 990.8 hPa
Analyst: MEH
Technician: MK

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/22/2022

Sample #139036
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
Aldicarb 0.400 0.125 Pass/<LOD Abamectin**** 0.500 0.125 Pass/<LOD
Acephate 0.400 0.125 Pass/<LOD Acequinocyl 2.000 0.250 Pass/<LOD
Acetamiprid 0.200 0.125 Pass/<LOD Azoxystrobin 0.200 0.125 Pass/<LOD
Bifenazate 0.200 0.125 Pass/<LOD Bifenthrin 0.200 0.125 Pass/<LOD
Boscalid 0.400 0.125 Pass/<LOD Carbaryl 0.200 0.125 Pass/<LOD
Carbofuran 0.200 0.125 Pass/<LOD Chlorantraniliprole 0.200 0.125 Pass/<LOD
Chlorfenapyr 1.000 0.500 Pass/<LOD Chlorpyrifos 0.200 0.125 Pass/<LOD
Clofentezine 0.200 0.125 Pass/<LOD Cyfluthrin** 1.000 0.500 Pass/<LOD
Cypermethrin*** 1.000 0.500 Pass/<LOD Daminozide 1.000 0.500 Pass/<LOD
DDVP (Dichlorvos) 1.000 0.250 Pass/<LOD Diazinon 0.200 0.125 Pass/<LOD
Dimethoate 0.200 0.125 Pass/<LOD Ethoprophos 0.200 0.125 Pass/<LOD
Etofenprox 0.400 0.125 Pass/<LOD Etoxazole 0.200 0.125 Pass/<LOD
Fenoxycarb 0.200 0.125 Pass/<LOD Fenpyroximate 0.400 0.125 Pass/<LOD
Fipronil 0.400 0.125 Pass/<LOD Flonicamid 1.000 0.125 Pass/<LOD
Fludioxonil 0.400 0.125 Pass/<LOD Hexythiazox 1.000 0.125 Pass/<LOD
Imazalil 0.200 0.125 Pass/<LOD Imidacloprid 0.400 0.125 Pass/<LOD
Kresoxim Methyl 0.400 0.125 Pass/<LOD Malathion 0.200 0.125 Pass/<LOD
Metalaxyl 0.200 0.125 Pass/<LOD Methiocarb 0.200 0.125 Pass/<LOD
Methomyl 0.400 0.125 Pass/<LOD Methyl Parathion 0.200 0.125 Pass/<LOD
MGK-264‡ 0.200 0.125 Pass/<LOD Myclobutanil 0.200 0.125 Pass/<LOD
Naled 0.500 0.125 Pass/<LOD Oxamyl 1.000 0.125 Pass/<LOD
Paclobutrazol 0.400 0.125 Pass/<LOD Permethrins† 0.200 0.125 Pass/<LOD
Phosmet 0.200 0.125 Pass/<LOD Piperonyl Butoxide 100.000 1.900 Not Tested
Prallethrin 0.200 0.125 Pass/<LOD Propiconazole 0.400 0.125 Pass/<LOD
Propoxur 0.200 0.125 Pass/<LOD Pyrethrins* 1.000 0.125 Pass/<LOD
Pyridaben 0.200 0.125 Pass/<LOD Spinosad***** 0.200 0.125 Pass/<LOD
Spiromesifen 0.200 0.125 Pass/<LOD Spirotetramat 0.200 0.125 Pass/<LOD
Spiroxamine‡ 0.400 0.125 Pass/<LOD Tebuconazole 0.400 0.125 Pass/<LOD
Thiacloprid 0.200 0.125 Pass/<LOD Thiamethoxam 0.200 0.125 Pass/<LOD
Trifloxystrobin 0.200 0.125 Pass/<LOD

* Pyrethrins are reported as the sum of Jasmolin I, Cinerin I, and Pyrethrin I
** Cyfluthrins are reported as the sum of isomers Cyfluthrin I, II, III, and IV
*** Cypermethrins are reported as the sum of isomers Cypermethrin I, II, III, and IV
**** Abamectin is reported as the sum of Avermectin B1a and Avermectin B1b
***** Spinosad is reported as the sum of Spinosyn A and Spinosyn D
† Permethrin and Prallethrin are reported as the sum of cis and trans isomers
‡ MGK-264 and Spiroximine are reported as the sum of isomers I and II
MRL - Maximum Residue Limit; LOD - Limit of Detection

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Pesticide QC Report

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Prepsheet ID#: MIPS221220a
Scale: XS205-MI1
Temp: 22.4 °C
Baro PE: 990.8 hPa
Analyst: MEH
Technician: MK

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/22/2022

Sample #139036
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS recovery% MSD recovery% Relative Percent Difference (%) QC Flag
Acephate N.D. 1 0.992 1.04 99.20 104.00 4.72
Acequinocyl N.D. 1 0.819 0.956 81.90 95.60 15.44
Acetamiprid N.D. 1 1.03 1.1 103.00 110.00 6.57
Aldicarb N.D. 1 1.01 0.974 101.00 97.40 3.63
Avermectin B1a N.D. 1 0.726 0.799 72.60 79.90 9.57
Azoxystrobin N.D. 1 0.719 0.768 71.90 76.80 6.59
Bifenazate N.D. 1 0.809 0.834 80.90 83.40 3.04
Bifenthrin N.D. 1 0.817 0.93 81.70 93.00 12.94
Boscalid N.D. 1 0.921 0.916 92.10 91.60 0.54
Carbaryl N.D. 1 1.05 0.991 105.00 99.10 5.78
Carbofuran N.D. 1 1.01 1.05 101.00 105.00 3.88
Chlorantraniliprole N.D. 1 0.76 0.689 76.00 68.90 9.80
Chlorfenapyr N.D. 1 0.978 1.16 97.80 116.00 17.03
Chlorpyriphos N.D. 1 0.808 0.863 80.80 86.30 6.58
Clofentezine N.D. 1 0.795 0.911 79.50 91.10 13.60
Cyfluthrin N.D. 1 0.838 0.88 83.80 88.00 4.89
Cypermethrin N.D. 1 0.718 0.732 71.80 73.20 1.93
Daminoside N.D. 1 0.3 0.352 30.00 35.20 15.95 LR
Diazanon N.D. 1 1.24 1.41 124.00 141.00 12.83 HR
Dichlorvos N.D. 1 1.11 1.09 111.00 109.00 1.82
Dimethoate N.D. 1 1.03 1.09 103.00 109.00 5.66
Ethoprophos N.D. 1 0.964 0.975 96.40 97.50 1.13
Etofenprox N.D. 1 0.901 0.991 90.10 99.10 9.51
Etoxazole N.D. 1 0.923 0.996 92.30 99.60 7.61
Fenoxycarb N.D. 1 0.923 0.958 92.30 95.80 3.72
Fenpyroximate N.D. 1 0.884 0.892 88.40 89.20 0.90
Fipronil N.D. 1 0.924 0.967 92.40 96.70 4.55
Flonicamid N.D. 1 1.04 1.07 104.00 107.00 2.84
Fludioxonil N.D. 1 1.01 1.07 101.00 107.00 5.77
Hexythiazox N.D. 1 0.813 0.87 81.30 87.00 6.77
Imazalil N.D. 1 0.879 0.904 87.90 90.40 2.80
Imidacloprid N.D. 1 0.944 1 94.40 100.00 5.76
Kresoxim-methyl N.D. 1 0.914 0.947 91.40 94.70 3.55
Malathion N.D. 1 0.822 0.838 82.20 83.80 1.93
Metalaxyl N.D. 1 0.881 0.883 88.10 88.30 0.23
Methiocarb N.D. 1 1.09 1.1 109.00 110.00 0.91
Methomyl N.D. 1 0.98 1.01 98.00 101.00 3.02
MGK-264 N.D. 1 0.938 0.938 93.80 93.80 0.00
Myclobutanil N.D. 1 0.911 0.927 91.10 92.70 1.74
Naled (dibrom) N.D. 1 0.637 0.639 63.70 63.90 0.31
Oxamyl N.D. 1 1.07 1.11 107.00 111.00 3.67
Paclobutrazol N.D. 1 0.907 0.939 90.70 93.90 3.47
Parathion-methyl N.D. 1 1.11 1.15 111.00 115.00 3.54
Permethrins N.D. 1 0.761 0.787 76.10 78.70 3.36
Phosmet N.D. 1 0.914 0.935 91.40 93.50 2.27
Piperonyl butoxide N/A N/A N/A N/A nan nan N/A
Prallethrin N.D. 1 0.857 0.861 85.70 86.10 0.47
Propiconazole N.D. 1 0.846 0.894 84.60 89.40 5.52
Propoxur N.D. 1 0.988 1.02 98.80 102.00 3.19
Pyrethrin N.D. 1 0.82 0.884 82.00 88.40 7.51
Pyridaben N.D. 1 0.899 0.985 89.90 98.50 9.13
SpinosynA N.D. 0.82 0.517 0.522 63.05 63.66 0.96
SpinosynD N.D. 0.18 0.158 0.153 87.78 85.00 3.22
Spiromesifen N.D. 1 0.652 0.679 65.20 67.90 4.06
Spirotetramat N.D. 1 0.938 1.07 93.80 107.00 13.15
Spiroxamine N.D. 1 0.979 1 97.90 100.00 2.12
Tebuconazole N.D. 1 0.948 0.929 94.80 92.90 2.02
Thiacloprid N.D. 1 0.939 0.941 93.90 94.10 0.21
Thiamethoxam N.D. 1 0.975 1 97.50 100.00 2.53
Trifloxystrobin N.D. 1 0.872 0.899 87.20 89.90 3.05

N.D. = Not Detected

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 30%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (60-120%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 3 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Microbiology Analysis

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Scale: XS205-MI1
Temp: 21.2 °C
Baro Pressure: 982.1 hPa
Analyst: SKM
Technician: SKM

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/23/2022

Sample #139036
Test Method MCFU (CFU/g) LOD (CFU/g) Status (CFU/g)
Total Aerobic Not Tested N/A 0 Not Tested
Yeast & Mold FE-66 (MPN) 100,001 100 Pass/409
Bile-tolerant gram-negative bacteria Not Tested N/A 0 Not Tested
Total Coliforms FE-66 (MPN) 1,001 100 Pass/409
E. Coli FE-65 (PCR) 1 1 Pass/Not Present
Salmonella FE-65 (PCR) 1 1 Pass/Not Present
Aspergillus spp. FE-65 (PCR) 1 1 Pass/Not Present

E. coli test screens exclusively for Shiga toxin-producing E. coli. TNTC values for yeast and mold are >150,000 CFU/g. TNTC values for total coliforms are >15,000 CFU/g. The maximum relative uncertainty is 6% for PCR and 40% for MPN.

TNTC = Too Numerous To Count. Maximum Colony Forming Units (MCFU) represents the minimum value that a test would need to achieve in order to be considered a failed result (if applicable). LOD is the Limit of Detection for the method. CFU/g = Colony Forming Units per gram. Aspergillus spp. consists of A. flavus, A. fumigatus, A. niger, and A. terreus.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 4 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Bio Analysis by
Microscopic Visual Inspection

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE02 Microscope Dissection
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Analyst: ARH
Technician: ARH

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/20/2022

Sample #139036
Sample #139036 Sample #139036
Severity Rating
Mold and Mildew 0 Pass
Mites (not Russet) 0 Pass
Russet Mites 0 Pass
Thrips 0 Pass
Aphids 0 Pass
Other Pests 0 Pass
Other Organic Matter 0 Pass
Inorganic Matter 0 Pass
Total Organic Foreign Matter <2% Pass
Inorganic Foreign Matter 0% Pass
Stem Material <5% Pass
Severity Rating Key
Correlating %FM ≤2% FM >2% FM
Severity Rating 0 None 1 Mild 2 Moderate 3 Severe Units
Mold and Mildew <2 2 3 4+ Locations/g
Mites (not russet) <1 1-5 6-10 11+ Amount/g
Russet Mites <1 1-10 11-50 51+ Amount/g
Thrips <1 1-3 4-5 6+ Amount/g
Aphids <1 1 2-4 5+ Amount/g
Other Pests <1 1-5 6-10 11+ Amount/g
Other Organic Matter <1 1-5 6-10 11+ Amount/g
Inorganic Matter 0 >0 Amount/g

In compliance with the MMFLA regulation, the action limits for foreign matter are as follows; no more that 2% organic matter, 5% stem material, and no presence of inorganic matter. Percent foreign contamination is based off of total surface area.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 5 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Heavy Metals Analysis

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE60 ICPMS-MI
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Prepsheet ID#: MIICP 221220fMK
Temp: 22.4 °C
Baro PE: 990.8 hPa
Analyst: MK
Technician: MK

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/20/2022

Sample #139036
Sample #139036 Sample #139036
Test Compound MRL (PPM) LOD (PPM) Status (PPM)
Arsenic 0.4 0.1 Pass/<LOD
Copper 0 1.5 Not Tested
Cadmium 0.4 0.1 Pass/0.147
Chromium 1.2 0.3 Pass/<LOD
Lead 1 0.25 Pass/<LOD
Mercury 0.2 0.05 Pass/<LOD
Nickel 1 0.25 Pass/<LOD

MRL - Maximum Residue Limit; LOD - Limit of Detection

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 6 of 6

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 139036-001
Client Name, Sample Details

Greenroom Gardens LLC AU-G-C-000910
Muskegon, MI 49441
Sample: Kush Mints
License: AU-G-C-000910
Type: Flower
Method: FE60 ICPMS-MI
Metrc Test Pkg#: 1A4050300032E11000000085
Metrc Source Pkg#: 1A4050300032E11000000082

Test Conditions

Prepsheet ID#: MIICP 221220fMK
Temp: 22.4 °C
Baro PE: 990.8 hPa
Analyst: MK
Technician: MK

Sample ID#: 139036

Harvest/Process Date: 12/19/2022
Serving Size (g): 1
Date Received: 12/19/2022
Test Date: 12/20/2022
Valid Through: 12/20/2023
Report Issued: 12/20/2022

Sample #139036
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD QC Flag
Arsenic N/D 0.160 0.152 95.00 0.142 88.75
Cadmium N/D 0.160 0.145 90.63 0.146 91.25
Chromium N/D 0.480 0.470 97.92 0.460 95.83
Copper N/D 2.400 2.666 111.08 2.695 112.29
Lead N/D 0.400 0.372 93.00 0.372 93.00
Mercury N/D 0.080 0.077 96.25 0.074 92.50
Nickel N/D 0.400 0.468 117.00 0.443 110.75

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

MEH QM Iron laboratories MEH QM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 7 of 6