IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 132203-001
Client Name, Sample Details

Holistic Healing Tim/RoseAnn
Troy, MI 48083
Sample: 1500 FS
Type: Generic Concentrate
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP210806B
Scale: XS205-MI2
Temp: 20.4 °C
Baro Pressure: 980.7 hPa
Analyst: VU
Technician: VU

Sample ID#: 132203

Harvest/Process Date: 08/03/2021
Serving Size (g): 0.92
Date Received: 08/03/2021
Test Date: 08/06/2021
Valid Through: 08/07/2022
Report Issued: 08/09/2021

Sample #132203
Sample #132203 Sample #132203
Test
Compounds
THC THCA CBD CBDA CBN CBG* CBC* THCV* CBDV* Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.2 N/D 6.3 0.0 N/D 0.5 0.2 N/D 0.1 7.4 0.2 6.4 7.4
Amount (mg/g) 2.3 N/D 63.2 0.4 N/D 5.3 1.6 N/D 1.0 73.8 2.3 63.6 73.8
Amount per Serving (mg) 2.1 N\D 58.1 0.4 N\D 4.9 1.5 N\D 0.9 67.9 Serving Size~ (g): 0.9
LOQ (mg/g) 0.29 0.29 0.29 0.29 0.29 0.29 0.29 0.29 0.29   %Decarb. THC CBD
±%RPD 0.97 1.29 0.18 0.54 0.21 0.88 1.15 1.55 0.53 100% 99

Serving size = 1 mL

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009. FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products. Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refer to the percentage of THC or CBD relative to THCA or CBDA, respectively.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 1

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 132203-001
Client Name, Sample Details

Holistic Healing Tim/RoseAnn
Troy, MI 48083
Sample: 1500 FS
Type: Generic Concentrate
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP210806B
Scale: XS205-MI2
Temp: 20.4 °C
Baro Pressure: 980.7 hPa
Analyst: VU
Technician: VU

Sample ID#: 132203

Harvest/Process Date: 08/03/2021
Serving Size (g): 0.92
Date Received: 08/03/2021
Test Date: 08/06/2021
Valid Through: 08/07/2022
Report Issued: 08/09/2021

Sample #132203
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) 0 5.37731 5.34622 99.42 5.31786 98.89 0.53
Cannabidiolic Acid (CBDA) 0 4.46479 4.49279 100.63 4.51721 101.17 0.54
Cannabigerol (CBG) 0 5.3459 5.40401 101.09 5.45203 101.99 0.88
Cannabidiol (CBD) 0 5.23303 5.31806 101.62 5.30853 101.44 0.18
Δ9-Tetrahydrocannabivarin (THCV) 0 1.77814 1.76119 99.05 1.73411 97.52 1.55
Cannabinol (CBN) 0 5.19841 5.18833 99.81 5.19928 100.02 0.21
Δ9-Tetrahydrocannabinol (THC) 0 5.29303 5.29401 100.02 5.24265 99.05 0.97
Cannabichromene (CBC) 0 5.29948 5.34699 100.90 5.28577 99.74 1.15
Tetrahydrocannabinolic acid (THCA) 0 4.9347 4.97739 100.87 4.91342 99.57 1.29

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 1