IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 128998-001
Client Name, Sample Details

JT Bills
Garden City, MI 48135
Sample: Cookies N Cream
Type: Flower
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP210203
Scale: XS205-MI2
Temp: 20.6 °C
Baro Pressure: 979.2 hPa
Analyst: KEB
Technician: ANJ

Sample ID#: 128998

Harvest/Process Date: 02/02/2021
Serving Size (g): 1
Date Received: 02/02/2021
Test Date: 02/03/2021
Valid Through: 02/04/2022
Report Issued: 02/04/2021

Sample #128998
Sample #128998 Sample #128998
Test
Compounds
THC THCA CBD CBDA CBN CBG* CBC* THCV* CBDV* Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.5 28.1 N/D 1.5 N/D 0.2 N/D N/D N/D 30.3 25.2 1.3 26.7
Amount (mg/g) 5.4 281.2 N/D 14.9 N/D 1.6 N/D N/D N/D 303.1 252.0 13.1 266.7
Amount per Serving (mg) 5.4 281.2 N\D 14.9 N\D 1.6 N\D N\D N\D 303.1 Serving Size~ (g): 1.0
LOQ (mg/g) 0.98 0.98 0.98 0.98 0.98 0.98 0.98 0.98 0.98   %Decarb. THC CBD
±%RPD 3.59 4.08 2.43 2.55 2.90 2.94 2.32 1.48 2.77 2 0%

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009. FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products. Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refer to the percentage of THC or CBD relative to THCA or CBDA, respectively.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 1

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 128998-001
Client Name, Sample Details

JT Bills
Garden City, MI 48135
Sample: Cookies N Cream
Type: Flower
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP210203
Scale: XS205-MI2
Temp: 20.6 °C
Baro Pressure: 979.2 hPa
Analyst: KEB
Technician: ANJ

Sample ID#: 128998

Harvest/Process Date: 02/02/2021
Serving Size (g): 1
Date Received: 02/02/2021
Test Date: 02/03/2021
Valid Through: 02/04/2022
Report Issued: 02/04/2021

Sample #128998
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) 0 5.85901 6.06284 103.48 6.23291 106.38 2.77
Cannabidiolic Acid (CBDA) 0 5.5991 5.80558 103.69 5.95574 106.37 2.55
Cannabigerol (CBG) 0 5.92018 6.12408 103.44 6.30691 106.53 2.94
Cannabidiol (CBD) 0 6.0842 6.25502 102.81 6.40886 105.34 2.43
Δ9-Tetrahydrocannabivarin (THCV) 0 5.98147 6.21382 103.88 6.30627 105.43 1.48
Cannabinol (CBN) 0 5.87817 5.99768 102.03 6.17399 105.03 2.90
Δ9-Tetrahydrocannabinol (THC) 0 6.56602 6.52347 99.35 6.76221 102.99 3.59
Cannabichromene (CBC) 0 5.89896 6.16141 104.45 6.30578 106.90 2.32
Tetrahydrocannabinolic acid (THCA) 0 5.71429 5.8294 102.01 6.07237 106.27 4.08

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This sample has not undergone random sampling and has not been tested for compliant state, batch representative testing. These results should therefore be used for research and development or quality control purposes only. Results apply to the sample as received.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 1