ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Marijuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 4092-004 Test Certificate #: 120808-001
Client Name, Sample Details

A Vant Garden
Oakridge, OR 97463
Sample: G.G.#4
License: 020-10064037561
Type: Adult Use Flower
Method: FE04U HPLC-UV
Metrc Test Pkg#: 1A4010300012111000001790
Metrc Source Pkg#: 1A4010300012111000001765
***Water Activity: 0.488
***Moisture: 13.41%

Test Conditions

Prepsheet ID#: ORP200103a
Scale: XS205-OR1
Temp: 23.2 °C
Baro Pressure: 1002 hPa
Analyst: HRM
Technician: TMR

Sample ID#: 120808

Lot #: G.G.#4 12-19-19
Batch #: 120808
Batch Weight (g): 6790
Harvest/Process Date: 12/19/2019
Serving Size (g): 1
Date Received: 01/02/2020
Test Date: 01/03/2020

Sample #120808
Sample #120808 Sample #120808
Test
Compounds
THC THCA CBD CBDA CBN CBG CBC THCV* CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 1.56 28.08 N/D 0.64 0.09 0.13 0.07 N/D N/D 30.56 26.18 0.56 27.03
Amount (mg/g) 15.55 280.77 N/D 6.39 0.93 1.31 0.67 N/D N/D 305.62 261.79 5.60 270.30
Amount per Serving (mg) 15.55 280.77 N\D 6.39 0.93 1.31 0.67 N\D N\D 305.62 Serving Size~ (g): 1.00
LOQ (mg/g) 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5 0.5   %Decarb. THC CBD
±%RPD 3.05 16.14 +/-5% 13.63 5.50 8.50 49.15 +/-5% +/-5% 5 0

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009. FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products. Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refer to the percentage of THC or CBD relative to THCA or CBDA, respectively.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Himashi Mead Iron laboratories Himashi Mead

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 1 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant QC Report

Testing Accreditation #: 4092-004 Test Certificate #: 120808-001
Client Name, Sample Details

A Vant Garden
Oakridge, OR 97463
Sample: G.G.#4
License: 020-10064037561
Type: Adult Use Flower
Method: FE04U HPLC-UV
Metrc Test Pkg#: 1A4010300012111000001790
Metrc Source Pkg#: 1A4010300012111000001765
***Water Activity: 0.488
***Moisture: 13.41%

Test Conditions

Prepsheet ID#: ORP200103a
Scale: XS205-OR1
Temp: 23.2 °C
Baro Pressure: 1002 hPa
Analyst: HRM
Technician: TMR

Sample ID#: 120808

Lot #: G.G.#4 12-19-19
Batch #: 120808
Batch Weight (g): 6790
Harvest/Process Date: 12/19/2019
Serving Size (g): 1
Date Received: 01/02/2020
Test Date: 01/03/2020

Sample #120808
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) N.D. 1 0 0.00 0 0.00 0
Cannabidiolic Acid (CBDA) N.D. 11.9 10.6 89.08 12.15 102.10 13.63
Cannabigerol (CBG) N.D. 1.77 1.69 95.48 1.84 103.95 8.50
Cannabidiol (CBD) N.D. 1 0 0.00 0 0.00 0
Δ9-Tetrahydrocannabivarin (THCV) N.D. 1 0 0.00 0 0.00 0
Cannabinol (CBN) N.D. 1.44 0.56 38.89 0.53 36.81 5.50
Δ9-Tetrahydrocannabinol (THC) N.D. 30.46 30.72 100.85 31.67 103.97 3.05
Cannabichromene (CBC) N.D. 1.27 1.47 115.75 0.89 70.08 49.15
Tetrahydrocannabinolic acid (THCA) N.D. 158.55 137.87 86.96 162.07 102.22 16.14

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Himashi Mead Iron laboratories Himashi Mead

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Pesticide Analysis by
Mass Spectrometer

Testing Accreditation #: 4092-004 Test Certificate #: 120808-001
Client Name, Sample Details

A Vant Garden
Oakridge, OR 97463
Sample: G.G.#4
License: 020-10064037561
Type: Adult Use Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4010300012111000001790
Metrc Source Pkg#: 1A4010300012111000001765

Test Conditions

Prepsheet ID#: ORPS200107a
Scale: XS205-OR1
Temp: 21.3 °C
Baro PE: 996.9 hPa
Analyst: HRM
Technician: HRM

Sample ID#: 120808

Lot #: G.G.#4 12-19-19
Batch #: 120808
Batch Weight (g): 6790
Harvest/Process Date: 12/19/2019
Serving Size (g): 1
Date Received: 01/02/2020
Test Date: 01/07/2020

Sample #120808
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
Aldicarb 0.400 0.121 Pass/<LOD Abamectin**** 0.500 0.121 Pass/<LOD
Acephate 0.400 0.121 Pass/<LOD Acequinocyl 2.000 0.242 Pass/<LOD
Acetamiprid 0.200 0.121 Pass/<LOD Azoxystrobin 0.200 0.121 Pass/<LOD
Bifenazate 0.200 0.121 Pass/<LOD Bifenthrin 0.200 0.121 Pass/<LOD
Boscalid 0.400 0.121 Pass/<LOD Carbaryl 0.200 0.121 Pass/<LOD
Carbofuran 0.200 0.121 Pass/<LOD Chlorantraniliprole 0.200 0.121 Pass/<LOD
Chlorfenapyr 1.000 0.484 Pass/<LOD Chlorpyrifos 0.200 0.121 Pass/<LOD
Clofentezine 0.200 0.121 Pass/<LOD Cyfluthrin** 1.000 0.484 Pass/<LOD
Cypermethrin*** 1.000 0.484 Pass/<LOD Daminozide 1.000 0.121 Pass/<LOD
DDVP (Dichlorvos) 1.000 0.242 Pass/<LOD Diazinon 0.200 0.121 Pass/<LOD
Dimethoate 0.200 0.121 Pass/<LOD Ethoprophos 0.200 0.121 Pass/<LOD
Etofenprox 0.400 0.121 Pass/<LOD Etoxazole 0.200 0.121 Pass/<LOD
Fenoxycarb 0.200 0.121 Pass/<LOD Fenpyroximate 0.400 0.121 Pass/<LOD
Fipronil 0.400 0.121 Pass/<LOD Flonicamid 1.000 0.121 Pass/<LOD
Fludioxonil 0.400 0.121 Pass/<LOD Hexythiazox 1.000 0.121 Pass/<LOD
Imazalil 0.200 0.121 Pass/<LOD Imidacloprid 0.400 0.121 Pass/<LOD
Kresoxim Methyl 0.400 0.121 Pass/<LOD Malathion 0.200 0.121 Pass/<LOD
Metalaxyl 0.200 0.121 Pass/<LOD Methiocarb 0.200 0.121 Pass/<LOD
Methomyl 0.400 0.121 Pass/<LOD Methyl Parathion 0.200 0.121 Pass/<LOD
MGK-264‡ 0.200 0.121 Pass/<LOD Myclobutanil 0.000 0.121 Pass/<LOD
Naled 0.500 0.121 Pass/<LOD Oxamyl 1.000 0.121 Pass/<LOD
Paclobutrazol 0.400 0.121 Pass/<LOD Permethrins† 0.200 0.121 Pass/<LOD
Phosmet 0.200 0.121 Pass/<LOD Piperonyl Butoxide 2.000 0.121 Pass/<LOD
Prallethrin 0.200 0.121 Pass/<LOD Propiconazole 0.400 0.121 Pass/<LOD
Propoxur 0.200 0.121 Pass/<LOD Pyrethrins* 1.000 0.121 Pass/<LOD
Pyridaben 0.200 0.121 Pass/<LOD Spinosad***** 0.200 0.121 Pass/<LOD
Spiromesifen 0.200 0.121 Pass/<LOD Spirotetramat 0.200 0.121 Pass/<LOD
Spiroxamine‡ 0.400 0.121 Pass/<LOD Tebuconazole 0.400 0.121 Pass/<LOD
Thiacloprid 0.200 0.121 Pass/<LOD Thiamethoxam 0.200 0.121 Pass/<LOD
Trifloxystrobin 0.200 0.121 Pass/<LOD

* Pyrethrins are reported as the sum of Jasmolin I, Cinerin I, and Pyrethrin I
** Cyfluthrins are reported as the sum of isomers Cyfluthrin I, II, III, and IV
*** Cypermethrins are reported as the sum of isomers Cypermethrin I, II, III, and IV
**** Abamectin is reported as the sum of Avermectin B1a and Avermectin B1b
***** Spinosad is reported as the sum of Spinosyn A and Spinosyn D
† Permethrin and Prallethrin are reported as the sum of cis and trans isomers
‡ MGK-264 and Spiroximine are reported as the sum of isomers I and II
MRL - Maximum Residue Limit; LOD - Limit of Detection

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Pesticide QC Report

Testing Accreditation #: 4092-004 Test Certificate #: 120808-001
Client Name, Sample Details

A Vant Garden
Oakridge, OR 97463
Sample: G.G.#4
License: 020-10064037561
Type: Adult Use Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A4010300012111000001790
Metrc Source Pkg#: 1A4010300012111000001765

Test Conditions

Prepsheet ID#: ORPS200107a
Scale: XS205-OR1
Temp: 21.3 °C
Baro PE: 996.9 hPa
Analyst: HRM
Technician: HRM

Sample ID#: 120808

Lot #: G.G.#4 12-19-19
Batch #: 120808
Batch Weight (g): 6790
Harvest/Process Date: 12/19/2019
Serving Size (g): 1
Date Received: 01/02/2020
Test Date: 01/07/2020

Sample #120808
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS recovery% MSD recovery% Relative Percent Difference (%) QC Flag
Acephate N.D. 1 0.856 0.933 85.60 93.30 8.61
Acequinocyl N.D. 1 0.758 0.754 75.80 75.40 0.53
Acetamiprid N.D. 1 0.967 0.874 96.70 87.40 10.10
Aldicarb N.D. 1 0.873 0.943 87.30 94.30 7.71
Avermectin B1a N.D. 0.97 0.732 0.812 75.46 83.71 10.36
Azoxystrobin N.D. 1 0.826 0.762 82.60 76.20 8.06
Bifenazate N.D. 1 0.62 0.7 62.00 70.00 12.12 LR
Bifenthrin N.D. 1 0.674 0.788 67.40 78.80 15.60 LR
Boscalid N.D. 1 0.694 0.706 69.40 70.60 1.71 LR
Carbaryl N.D. 1 0.945 0.858 94.50 85.80 9.65
Carbofuran N.D. 1 0.814 0.736 81.40 73.60 10.06
Chlorantraniliprole N.D. 1 0.456 0.522 45.60 52.20 13.50 LR
Chlorfenapyr N.D. 1 0.829 0.868 82.90 86.80 4.60
Chlorpyriphos N.D. 1 0.738 0.878 73.80 87.80 17.33
Clofentezine N.D. 1 0.45 0.49 45.00 49.00 8.51 LR
Cyfluthrin N.D. 1 0.662 0.661 66.20 66.10 0.15
Cypermethrin N.D. 1 0.807 0.769 80.70 76.90 4.82
Daminoside N.D. 1 0.562 0.701 56.20 70.10 22.01 LRQ
Diazanon N.D. 1 3.97 4.77 397.00 477.00 18.31 HR
Dichlorvos N.D. 1 0.898 0.821 89.80 82.10 8.96 HR
Dimethoate N.D. 1 1.01 0.857 101.00 85.70 16.39
Ethoprophos N.D. 1 0.783 0.638 78.30 63.80 20.41
Etofenprox N.D. 1 0.803 0.746 80.30 74.60 7.36
Etoxazole N.D. 1 0.859 0.984 85.90 98.40 13.56
Fenoxycarb N.D. 1 0.454 0.548 45.40 54.80 18.76 LR
Fenpyroximate N.D. 1 0.883 0.92 88.30 92.00 4.10
Fipronil N.D. 1 1 1.06 100.00 106.00 5.83
Flonicamid N.D. 1 0.881 0.828 88.10 82.80 6.20
Fludioxonil N.D. 1 1.07 1.11 107.00 111.00 3.67
Hexythiazox N.D. 1 0.467 0.591 46.70 59.10 23.44 LRQ
Imazalil N.D. 1 0.552 0.595 55.20 59.50 7.50 LR
Imidacloprid N.D. 1 0.89 0.818 89.00 81.80 8.43
Kresoxim-methyl N.D. 1 0.448 0.59 44.80 59.00 27.36 LRQ
Malathion N.D. 1 0.575 0.701 57.50 70.10 19.75 LR
Metalaxyl N.D. 1 0.656 0.726 65.60 72.60 10.13 LR
Methiocarb N.D. 1 0.886 0.805 88.60 80.50 9.58
Methomyl N.D. 1 0.601 0.736 60.10 73.60 20.19 LRQ
MGK-264 N.D. 1 0.706 0.686 70.60 68.60 2.87
Myclobutanil N.D. 1 0.544 0.573 54.40 57.30 5.19 LR
Naled (dibrom) N.D. 1 0.742 0.883 74.20 88.30 17.35
Oxamyl N.D. 1 0.853 0.782 85.30 78.20 8.69
Paclobutrazol N.D. 1 0.479 0.538 47.90 53.80 11.60
Parathion-methyl N.D. 1 0.892 0.893 89.20 89.30 0.11
Permethrins N.D. 1 0.715 0.731 71.50 73.10 2.21
Phosmet N.D. 1 0.567 0.677 56.70 67.70 17.68 LR
Piperonyl butoxide N.D. 1 0.865 0.945 86.50 94.50 8.84
Prallethrin N.D. 1 0.536 0.614 53.60 61.40 13.57 LR
Propiconazole N.D. 1 0.486 0.553 48.60 55.30 12.90 LR
Propoxur N.D. 1 0.734 0.832 73.40 83.20 12.52
Pyrethrin N.D. 0.65 0.406 0.477 62.46 73.38 16.08 LR
Pyridaben N.D. 1 0.775 0.824 77.50 82.40 6.13
SpinosynA N.D. 0.84 0.396 0.575 47.14 68.45 36.87 LRQ
SpinosynD N.D. 0.16 0.0279 0.043 17.44 26.88 42.60 LRQ
Spiromesifen N.D. 1 1.03 1 103.00 100.00 2.96
Spirotetramat N.D. 1 0.836 0.973 83.60 97.30 15.15
Spiroxamine N.D. 1 0.556 0.622 55.60 62.20 11.21
Tebuconazole N.D. 1 0.428 0.555 42.80 55.50 25.84 LRQ
Thiacloprid N.D. 1 0.814 0.761 81.40 76.10 6.73
Thiamethoxam N.D. 1 0.547 0.638 54.70 63.80 15.36 LR
Trifloxystrobin N.D. 1 0.855 0.867 85.50 86.70 1.39

N.D. = Not Detected

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 30%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (60-120%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 3 of 3