ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Marijuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 4092-004 Test Certificate #: 119688-001
Client Name, Sample Details

Revel Farms
Eugene, OR 97425
Sample: Goji Berry
License: 10069798E0F
Type: Adult Use Flower
Method: FE04U HPLC-UV
Metrc Test Pkg#: 1A401030000F0A1000001110
Metrc Source Pkg#: 1A401030000F0A1000001108
***Water Activity: 0.51
***Moisture: 12.26%

Test Conditions

Prepsheet ID#: ORP191025a
Scale: XS205-OR1
Temp: 21.4 °C
Baro Pressure: 1006 hPa
Analyst: HRM
Technician: EDT

Sample ID#: 119688

Lot #: Goji Berry 09/23/19
Batch #: 119688
Batch Weight (g): 6810
Harvest/Process Date: 09/23/2019
Serving Size (g): 1
Date Received: 10/25/2019
Test Date: 10/25/2019

Sample #119688
Sample #119688 Sample #119688
Test
Compounds
THC THCA CBD CBDA CBN CBG CBC THCV* CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 1.06 16.64 N/D 0.84 0.05 0.07 N/D N/D N/D 18.67 15.66 0.74 16.52
Amount (mg/g) 10.64 166.42 N/D 8.43 0.52 0.71 N/D N/D N/D 186.72 156.59 7.39 165.21
Amount per Serving (mg) 10.64 166.42 N\D 8.43 0.52 0.71 N\D N\D N\D 186.72 Serving Size~ (g): 1.00
LOQ (mg/g) 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05 0.05   %Decarb. THC CBD
±%RPD 4.03 11.23 +/-5% 6.69 +/-5% +/-5% +/-5% +/-5% +/-5% 6 0

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009. FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products. Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refer to the percentage of THC or CBD relative to THCA or CBDA, respectively.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Himashi Mead Iron laboratories Himashi Mead

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 1 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant QC Report

Testing Accreditation #: 4092-004 Test Certificate #: 119688-001
Client Name, Sample Details

Revel Farms
Eugene, OR 97425
Sample: Goji Berry
License: 10069798E0F
Type: Adult Use Flower
Method: FE04U HPLC-UV
Metrc Test Pkg#: 1A401030000F0A1000001110
Metrc Source Pkg#: 1A401030000F0A1000001108
***Water Activity: 0.51
***Moisture: 12.26%

Test Conditions

Prepsheet ID#: ORP191025a
Scale: XS205-OR1
Temp: 21.4 °C
Baro Pressure: 1006 hPa
Analyst: HRM
Technician: EDT

Sample ID#: 119688

Lot #: Goji Berry 09/23/19
Batch #: 119688
Batch Weight (g): 6810
Harvest/Process Date: 09/23/2019
Serving Size (g): 1
Date Received: 10/25/2019
Test Date: 10/25/2019

Sample #119688
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) N.D. 1 0 0.00 0 0.00 0
Cannabidiolic Acid (CBDA) N.D. 1.29 1.39 107.75 1.3 100.78 6.69
Cannabigerol (CBG) N.D. 0.16 0.17 106.25 0.17 106.25 0.00
Cannabidiol (CBD) N.D. 1 0 0.00 0 0.00 0
Δ9-Tetrahydrocannabivarin (THCV) N.D. 1 0 0.00 0 0.00 0
Cannabinol (CBN) N.D. 0.1 0.06 60.00 0.06 60.00 0.00
Δ9-Tetrahydrocannabinol (THC) N.D. 3.31 3.16 95.47 3.29 99.40 4.03
Cannabichromene (CBC) N.D. 0.08 0.11 137.50 0.11 137.50 0.00
Tetrahydrocannabinolic acid (THCA) N.D. 18.31 18.72 102.24 16.73 91.37 11.23

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Himashi Mead Iron laboratories Himashi Mead

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Pesticide Analysis by
Mass Spectrometer

Testing Accreditation #: 4092-004 Test Certificate #: 119688-001
Client Name, Sample Details

Revel Farms
Eugene, OR 97425
Sample: Goji Berry
License: 10069798E0F
Type: Adult Use Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A401030000F0A1000001110
Metrc Source Pkg#: 1A401030000F0A1000001108

Test Conditions

Prepsheet ID#: ORPS191031a
Scale: Veritas-OR1
Temp: 22 °C
Baro PE: 1011 hPa
Analyst: JER
Technician: HRM

Sample ID#: 119688

Lot #: Goji Berry 09/23/19
Batch #: 119688
Batch Weight (g): 6810
Harvest/Process Date: 09/23/2019
Serving Size (g): 1
Date Received: 10/25/2019
Test Date: 11/01/2019

Sample #119688
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
Aldicarb 0.400 0.122 Pass/<LOD Abamectin**** 0.500 0.122 Pass/<LOD
Acephate 0.400 0.122 Pass/<LOD Acequinocyl 2.000 0.244 Pass/<LOD
Acetamiprid 0.200 0.122 Pass/<LOD Azoxystrobin 0.200 0.122 Pass/<LOD
Bifenazate 0.200 0.122 Pass/<LOD Bifenthrin 0.200 0.122 Pass/<LOD
Boscalid 0.400 0.122 Pass/<LOD Carbaryl 0.200 0.122 Pass/<LOD
Carbofuran 0.200 0.122 Pass/<LOD Chlorantraniliprole 0.200 0.122 Pass/<LOD
Chlorfenapyr 1.000 0.489 Pass/<LOD Chlorpyrifos 0.200 0.122 Pass/<LOD
Clofentezine 0.200 0.122 Pass/<LOD Cyfluthrin** 1.000 0.489 Pass/<LOD
Cypermethrin*** 1.000 0.489 Pass/<LOD Daminozide 1.000 0.122 Pass/<LOD
DDVP (Dichlorvos) 1.000 0.244 Pass/<LOD Diazinon 0.200 0.122 Pass/<LOD
Dimethoate 0.200 0.122 Pass/<LOD Ethoprophos 0.200 0.122 Pass/<LOD
Etofenprox 0.400 0.122 Pass/<LOD Etoxazole 0.200 0.122 Pass/<LOD
Fenoxycarb 0.200 0.122 Pass/<LOD Fenpyroximate 0.400 0.122 Pass/<LOD
Fipronil 0.400 0.122 Pass/<LOD Flonicamid 1.000 0.122 Pass/<LOD
Fludioxonil 0.400 0.122 Pass/<LOD Hexythiazox 1.000 0.122 Pass/<LOD
Imazalil 0.200 0.122 Pass/<LOD Imidacloprid 0.400 0.122 Pass/<LOD
Kresoxim Methyl 0.400 0.122 Pass/<LOD Malathion 0.200 0.122 Pass/<LOD
Metalaxyl 0.200 0.122 Pass/<LOD Methiocarb 0.200 0.122 Pass/<LOD
Methomyl 0.400 0.122 Pass/<LOD Methyl Parathion 0.200 0.122 Pass/<LOD
MGK-264‡ 0.200 0.122 Pass/<LOD Myclobutanil 0.000 0.122 Pass/<LOD
Naled 0.500 0.122 Pass/<LOD Oxamyl 1.000 0.122 Pass/<LOD
Paclobutrazol 0.400 0.122 Pass/<LOD Permethrins† 0.200 0.122 Pass/<LOD
Phosmet 0.200 0.122 Pass/<LOD Piperonyl Butoxide 2.000 0.122 Pass/<LOD
Prallethrin 0.200 0.122 Pass/<LOD Propiconazole 0.400 0.122 Pass/<LOD
Propoxur 0.200 0.122 Pass/<LOD Pyrethrins* 1.000 0.122 Pass/<LOD
Pyridaben 0.200 0.122 Pass/<LOD Spinosad***** 0.200 0.122 Pass/<LOD
Spiromesifen 0.200 0.122 Pass/<LOD Spirotetramat 0.200 0.122 Pass/<LOD
Spiroxamine‡ 0.400 0.122 Pass/<LOD Tebuconazole 0.400 0.122 Pass/<LOD
Thiacloprid 0.200 0.122 Pass/<LOD Thiamethoxam 0.200 0.122 Pass/<LOD
Trifloxystrobin 0.200 0.122 Pass/<LOD

* Pyrethrins are reported as the sum of Jasmolin I, Cinerin I, and Pyrethrin I
** Cyfluthrins are reported as the sum of isomers Cyfluthrin I, II, III, and IV
*** Cypermethrins are reported as the sum of isomers Cypermethrin I, II, III, and IV
**** Abamectin is reported as the sum of Avermectin B1a and Avermectin B1b
***** Spinosad is reported as the sum of Spinosyn A and Spinosyn D
† Permethrin and Prallethrin are reported as the sum of cis and trans isomers
‡ MGK-264 and Spiroximine are reported as the sum of isomers I and II
MRL - Maximum Residue Limit; LOD - Limit of Detection

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 3

ORELAP     IRON Laboratories     NELAP Accredited Laboratory

ORELAP Cert No. 4092-004
OLCC No. 1002158CD2E

Oregon Compliant Pesticide QC Report

Testing Accreditation #: 4092-004 Test Certificate #: 119688-001
Client Name, Sample Details

Revel Farms
Eugene, OR 97425
Sample: Goji Berry
License: 10069798E0F
Type: Adult Use Flower
Method: FE-52 (EN 15662 & AOAC 2007.01)
Metrc Test Pkg#: 1A401030000F0A1000001110
Metrc Source Pkg#: 1A401030000F0A1000001108

Test Conditions

Prepsheet ID#: ORPS191031a
Scale: Veritas-OR1
Temp: 22 °C
Baro PE: 1011 hPa
Analyst: JER
Technician: HRM

Sample ID#: 119688

Lot #: Goji Berry 09/23/19
Batch #: 119688
Batch Weight (g): 6810
Harvest/Process Date: 09/23/2019
Serving Size (g): 1
Date Received: 10/25/2019
Test Date: 11/01/2019

Sample #119688
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS recovery% MSD recovery% Relative Percent Difference (%) QC Flag
Acephate N.D. 1 1.19 1.16 119.00 116.00 2.55
Acequinocyl N.D. 1 0.977 0.834 97.70 83.40 15.79
Acetamiprid N.D. 1 1.29 1.29 129.00 129.00 0.00
Aldicarb N.D. 1 1.24 1.19 124.00 119.00 4.12
Avermectin B1a N.D. 0.97 1.27 1.29 130.93 132.99 1.56 HR
Azoxystrobin N.D. 1 1.19 1.3 119.00 130.00 8.84
Bifenazate N.D. 1 0.85 0.841 85.00 84.10 1.06
Bifenthrin N.D. 1 1.23 1.26 123.00 126.00 2.41
Boscalid N.D. 1 1.22 1.11 122.00 111.00 9.44
Carbaryl N.D. 1 1.26 1.22 126.00 122.00 3.23
Carbofuran N.D. 1 1.34 1.41 134.00 141.00 5.09 HR
Chlorantraniliprole N.D. 1 1.17 1.23 117.00 123.00 5.00
Chlorfenapyr N.D. 1 0.789 0.823 78.90 82.30 4.22
Chlorpyriphos N.D. 1 1.1 1.14 110.00 114.00 3.57
Clofentezine N.D. 1 1.08 1.09 108.00 109.00 0.92
Cyfluthrin N.D. 1 1.44 1.57 144.00 157.00 8.64 HR
Cypermethrin N.D. 1 1.56 1.44 156.00 144.00 8.00 HR
Daminoside N.D. 1 0.415 0.39 41.50 39.00 6.21 LR
Diazanon N.D. 1 4.34 4.29 434.00 429.00 1.16 HR
Dichlorvos N.D. 1 1.25 1.25 125.00 125.00 0.00
Dimethoate N.D. 1 1.35 1.36 135.00 136.00 0.74 HR
Ethoprophos N.D. 1 1.22 1.21 122.00 121.00 0.82
Etofenprox N.D. 1 1.22 1.24 122.00 124.00 1.63
Etoxazole N.D. 1 1.37 1.23 137.00 123.00 10.77 HR
Fenoxycarb N.D. 1 1.18 1.08 118.00 108.00 8.85
Fenpyroximate N.D. 1 1.45 1.38 145.00 138.00 4.95 HR
Fipronil N.D. 1 1.14 1.2 114.00 120.00 5.13
Flonicamid N.D. 1 1.5 1.5 150.00 150.00 0.00 HR
Fludioxonil N.D. 1 1.26 1.31 126.00 131.00 3.89 HR
Hexythiazox N.D. 1 0.942 1.05 94.20 105.00 10.84
Imazalil N.D. 1 1.21 1.19 121.00 119.00 1.67
Imidacloprid N.D. 1 1.42 1.44 142.00 144.00 1.40 HR
Kresoxim-methyl N.D. 1 1.02 1.03 102.00 103.00 0.98
Malathion N.D. 1 1.19 1.14 119.00 114.00 4.29
Metalaxyl N.D. 1 1.33 1.35 133.00 135.00 1.49 HR
Methiocarb N.D. 1 1.35 1.3 135.00 130.00 3.77 HR
Methomyl N.D. 1 1.3 1.2 130.00 120.00 8.00
MGK-264 N.D. 1 1.5 1.4 150.00 140.00 6.90 HR
Myclobutanil N.D. 1 1.38 1.22 138.00 122.00 12.31 HR
Naled (dibrom) N.D. 1 0.41 0.451 41.00 45.10 9.52 LR
Oxamyl N.D. 1 1.45 1.45 145.00 145.00 0.00 HR
Paclobutrazol N.D. 1 1.23 1.29 123.00 129.00 4.76
Parathion-methyl N.D. 1 1.09 1.05 109.00 105.00 3.74
Permethrins N.D. 1 1.32 1.32 132.00 132.00 0.00 HR
Phosmet N.D. 1 1.13 1.14 113.00 114.00 0.88
Piperonyl butoxide N.D. 1 1.15 1.17 115.00 117.00 1.72
Prallethrin N.D. 1 1.17 1.14 117.00 114.00 2.60
Propiconazole N.D. 1 1.08 0.998 108.00 99.80 7.89
Propoxur N.D. 1 1.21 1.19 121.00 119.00 1.67
Pyrethrin N.D. 0.65 0.803 0.708 123.54 108.92 12.57
Pyridaben N.D. 1 1.14 1.15 114.00 115.00 0.87
SpinosynA N.D. 0.84 0.547 0.776 65.12 92.38 34.62 LR
SpinosynD N.D. 0.16 0.129 0.158 80.63 98.75 20.21 Q
Spiromesifen N.D. 1 0.197 0.18 19.70 18.00 9.02 LR
Spirotetramat N.D. 1 0.922 0.918 92.20 91.80 0.43
Spiroxamine N.D. 1 1.32 1.34 132.00 134.00 1.50 HR
Tebuconazole N.D. 1 1.2 1.19 120.00 119.00 0.84
Thiacloprid N.D. 1 1.42 1.43 142.00 143.00 0.70 HR
Thiamethoxam N.D. 1 1.31 1.33 131.00 133.00 1.52 HR
Trifloxystrobin N.D. 1 1.23 1.23 123.00 123.00 0.00

N.D. = Not Detected

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 30%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (60-120%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

SOP FE-01-OR9 was used in accordance with OAR 333-007 for sampling. All marijuana items are sampled and tested in accordance with OAR 333-007-0300 to 333-007-490 and OAR 333-064-0100 to 333-064-0110.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 3 of 3