IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 116736-001
Client Name, Sample Details

Gabe Szakal
Southfield, MI 48076
Sample: Bee Cookies
Type: Flower
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP190502
Scale: XS205-MI2
Temp: 23.7 °C
Baro Pressure: 984.3 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 116736

Harvest/Process Date: 05/02/2019
Serving Size (g): 1
Date Received: 05/02/2019
Test Date: 05/02/2019
Valid Through: 05/02/2020
Report Issued: 05/03/2019

Sample #116736
Sample #116736 Sample #116736
Test
Compounds
THC THCA CBD CBDA CBN CBG* CBC* THCV* CBDV* Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 0.7 19.1 N/D 1.3 N/D N/D N/D N/D N/D 21.1 17.5 1.1 18.6
Amount (mg/g) 7.1 191.0 N/D 12.8 N/D N/D N/D N/D N/D 210.9 174.6 11.2 185.8
Amount per Serving (mg) 7.1 191.0 N\D 12.8 N\D N\D N\D N\D N\D 210.9 Serving Size~ (g): 1.0
LOQ (mg/g) 0.96 0.96 0.96 0.96 0.96 0.96 0.96 0.96 0.96   %Decarb. THC CBD
±%RPD 2.23 2.71 0.68 4.27 2.48 0.35 5.10 5.73 6.85 4 0%

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009. FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products. Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refer to the percentage of THC or CBD relative to THCA or CBDA, respectively.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 2

IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 116736-001
Client Name, Sample Details

Gabe Szakal
Southfield, MI 48076
Sample: Bee Cookies
Type: Flower
Method: FE04U HPLC-UV

Test Conditions

Prepsheet ID#: MIP190502
Scale: XS205-MI2
Temp: 23.7 °C
Baro Pressure: 984.3 hPa
Analyst: KEB
Technician: KEB

Sample ID#: 116736

Harvest/Process Date: 05/02/2019
Serving Size (g): 1
Date Received: 05/02/2019
Test Date: 05/02/2019
Valid Through: 05/02/2020
Report Issued: 05/03/2019

Sample #116736
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) 0 4.23019 4.88173 115.40 4.55833 107.76 6.85
Cannabidiolic Acid (CBDA) 0 4.63007 4.76618 102.94 4.56685 98.63 4.27
Cannabigerol (CBG) 0 4.55025 4.74954 104.38 4.76619 104.75 0.35
Cannabidiol (CBD) 0 4.77668 4.9814 104.29 5.01555 105.00 0.68
Δ9-Tetrahydrocannabivarin (THCV) 0 4.56473 4.7407 103.85 5.02054 109.99 5.73
Cannabinol (CBN) 0 4.34937 4.51413 103.79 4.6273 106.39 2.48
Δ9-Tetrahydrocannabinol (THC) 0 5.09018 4.88289 95.93 4.993 98.09 2.23
Cannabichromene (CBC) 0 4.6294 4.72754 102.12 4.97516 107.47 5.10
Tetrahydrocannabinolic acid (THCA) 0 4.50008 4.89563 108.79 4.76459 105.88 2.71

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 2

IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

Bio Analysis by
Microscopic Visual Inspection

Testing Accreditation #: 77802 Test Certificate #: 116736-001
Client Name, Sample Details

Gabe Szakal
Southfield, MI 48076
Sample: Bee Cookies
Type: Flower
Method: FE02 Microscope Dissection

Test Conditions

Analyst: ARH
Technician: SEB

Sample ID#: 116736

Harvest/Process Date: 05/02/2019
Serving Size (g): 1
Date Received: 05/02/2019
Test Date: 01/01/1970
Valid Through: 01/01/1970
Report Issued: 05/02/2019

Sample #116736
Sample #116736 Sample #116736
Impurities: Not Detected
Molds: No Types: N/A
Pests: No Types: N/A
Anthers: No
Detritus: No
Fibers: No
Hair: No
Mature Seeds: No
Premature Seeds: No
Unidentified Foliar Spray: No
Other Impurities: No
*Moisture -

The gland heads range from small to large in size and are abundant. Slightly more of them are clear than milky with a moderate amount turning amber in color. Some of the trichomes are damaged and degrading. A few of the trichomes are reduced due to stress. Less than 2% foreign matter. Less than 5% stem material. If you are interested in additional testing for quantitative microbial results, please call the lab within a 1-2 days to add on additional testing.

*Moisture is not part of the accredited scope of Iron Laboratories.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 2