IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 114858-001
Client Name, Sample Details

Plant Science Laboratories, LLC
Buffalo, NY 14215
Sample: Industrial Hemp Field Crop (Lot R4-AN-82980-AN)
Type: Flower
Method: FE04U HPLC1100-1

Test Conditions

Prepsheet ID#: MIP190205b
Scale: XS205-MI2
Temp: 22.9 °C
Baro PE: 983.7 hPa
Analyst: JRT
Technician: JRT

Sample ID#: 114858

Harvest/Process Date: 02/05/2019
Serving Size (g): 1
Date Received: 02/05/2019
Test Date: 02/05/2019
Valid Through: 02/05/2020

Sample #114858
Sample #114858 Sample #114858
Test
Compounds
THC THCA CBD CBDA CBN CBG* CBC* THCV* CBDV* Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) N/D 0.4 0.2 3.0 N/D N/D N/D N/D N/D 3.6 0.4 2.8 3.2
Amount (mg/g) N/D 4.2 2.1 29.5 N/D N/D N/D N/D N/D 35.8 3.7 28.0 31.7
Amount per Serving (mg) N\D 4.2 2.1 29.5 N\D N\D N\D N\D N\D 35.8 Serving Size~ (g): 1.0
LOQ (mg/g) 1.05 1.05 1.05 1.05 1.05 1.05 1.05 1.05 1.05   %Decarb. THC CBD
±%RPD 0.7 0.1 0.9 1.3 0.5 0.7 4.0 3.9 3.5 0% 7

Results not adjusted for moisture.

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45.

Total THC and CBD is the calculated sum of THC or CBD and the amount of THC or CBD derived from THCA or CBDA, respectively. These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value. Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, CBG, CBC, THCV, and CBDV.

%Decarb. THC and CBD refers to the percentage of THC or CBD relative to THCA or CBDA, respectively.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 2

IRON LaboratoriesPJLA

ISO/IEC 17025:2005 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 114858-001
Client Name, Sample Details

Plant Science Laboratories, LLC
Buffalo, NY 14215
Sample: Industrial Hemp Field Crop (Lot R4-AN-82980-AN)
Type: Flower
Method: FE04U HPLC1100-1

Test Conditions

Prepsheet ID#: MIP190205b
Scale: XS205-MI2
Temp: 22.9 °C
Baro PE: 983.7 hPa
Analyst: JRT
Technician: JRT

Sample ID#: 114858

Harvest/Process Date: 02/05/2019
Serving Size (g): 1
Date Received: 02/05/2019
Test Date: 02/05/2019
Valid Through: 02/05/2020

Sample #114858
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) 0 1.794113474 1.83158225 102.09 1.769035106 98.60 3.47%
Cannabidiolic Acid (CBDA) 0 5.155879596 5.11939326 99.29 5.053673349 98.02 1.29%
Cannabigerol (CBG) 0 3.854533693 3.729030535 96.74 3.754696784 97.41 0.69%
Cannabidiol (CBD) 0 5.205197293 5.185410817 99.62 5.231755869 100.51 0.89%
Δ9-Tetrahydrocannabivarin (THCV) 0 2.529547038 2.457316437 97.14 2.364042901 93.46 3.87%
Cannabinol (CBN) 0 5.040377863 4.932605931 97.86 4.956017803 98.33 0.47%
Δ9-Tetrahydrocannabinol (THC) 0 6.422072081 6.046940799 94.16 6.088897805 94.81 0.69%
Cannabichromene (CBC) 0 4.307002719 4.687875627 108.84 4.880368132 113.31 4.02%
Tetrahydrocannabinolic acid (THCA) 0 5.598105467 5.558211623 99.29 5.563218118 99.38 0.09%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

ACR - LM Iron laboratories ACR - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2005 Testing Laboratory laboratory, accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390