ORELAPIRON LaboratoriesNELAP Accredited Laboratory

ORELAP Cert No. 4092-002
OLCC No. 1002158CD2E

Marijuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 4092-002 Test Certificate #: 114134-001
Client Name, Sample Details

Modern Moonshine
Junction city, OR 97448
Sample: Delta 9 Distillate
Type: Extract
Method: FE04C-OR4 HPLC1100-OR1

Test Conditions

Scale: XS205-OR1
Temp: 22.3 °C
Baro Pressure: 1004 hPa
Analyst: HRM
Technician: EDT

Sample ID#: 114134

Lot #: R&D
Batch #: 114134
Harvest/Process Date: 01/01/2019
Serving Size (g): 0.1
Package Weight (g):1
Date Received: 01/02/2019
Test Date: 01/07/2019

Sample #114134
Sample #114134 Sample #114134
Test
Compounds
CBDV CBDA CBD CBC CBG CBN THC THCA THCV   Calc. Avg
Avg. Amount N/D N/D 2.33 mg/g 144.64 mg/g 42.77 mg/g 45.26 mg/g 749.51 mg/g 3.55 mg/g 5.18 mg/g CBD 2.33 mg/g
LOQ 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g 0.05 mg/g THC 752.62 mg/g
Amount per Serving~ N/D N/D 0.23 mg/srv 14.46 mg/srv 4.28 mg/srv 4.53 mg/srv 74.95 mg/srv 0.36 mg/srv 0.52 mg/srv Serving Size~ (g): 0.10 g
Uncertainty ±5% RPD ±5% RPD ±5% RPD ±5% RPD ±5% RPD ±5% RPD ±5% RPD ±5% RPD ±5% RPD

LOQ = Limit of Quantitation

RPD is relative percent difference between the LCS and LCS duplicate.

%RPD = Relative Percent Difference; Min. Value = Minimum Detectable Amount, CFU = Colony Forming Units, N/D = Not Detected

*** Designated tests that use the method FE-45.

Total CBD or THC is the calculated sum of CBD or THC and the amount of CBD or THC derived from CBDA or THCA, respectively. These values are calculated by adding a molar correction factor of 0.877 to the CBDA or THCA value.

Calc. Avg. is the calculated average of total CBD or total THC for all increments tested.

This sample has not been tested according to OAR 333-007. These results should therefore be used for research and development or quality control purposes only.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 1 of 3

ORELAPIRON LaboratoriesNELAP Accredited Laboratory

ORELAP Cert No. 4092-002
OLCC No. 1002158CD2E

Pesticide Analysis by
Mass Spectrometer

Testing Accreditation #: 4092-002 Test Certificate #: 114134-001
Client Name, Sample Details

Modern Moonshine
Junction city, OR 97448
Sample: Delta 9 Distillate
Type: Extract
Method: FE-52 (EN 15662 & AOAC 2007.01)

Test Conditions

Prepsheet ID#: ORPS190111a
Scale: Veritas-OR1
Temp: 22 °C
Baro PE: 995.8 hPa
Analyst: JDM
Technician: JDM

Sample ID#: 114134

Lot #: R&D
Batch #: 114134
Harvest/Process Date: 01/01/2019
Serving Size (g): 0.1
Package Weight (g):1
Date Received: 01/02/2019
Test Date: 01/12/2019

Sample #114134
Compound MRL (µg/g) LOD (µg/g) Status (µg/g)   Compound MRL (µg/g) LOD (µg/g) Status (µg/g)
Aldicarb 0.400 0.120 Pass/<LOD Abamectin**** 0.500 0.120 Pass/<LOD
Acephate 0.400 0.120 Pass/<LOD Acequinocyl 2.000 0.240 Pass/<LOD
Acetamiprid 0.200 0.120 Pass/<LOD Azoxystrobin 0.200 0.120 Pass/<LOD
Bifenazate 0.200 0.120 Pass/<LOD Bifenthrin 0.200 0.120 Pass/<LOD
Boscalid 0.400 0.120 Pass/<LOD Carbaryl 0.200 0.120 Pass/<LOD
Carbofuran 0.200 0.120 Pass/<LOD Chlorantraniliprole 0.200 0.120 Pass/<LOD
Chlorfenapyr 1.000 0.480 Pass/<LOD Chlorpyrifos 0.200 0.120 Pass/<LOD
Clofentezine 0.200 0.120 Pass/<LOD Cyfluthrin** 1.000 0.480 Pass/<LOD
Cypermethrin*** 1.000 0.480 Pass/<LOD Daminozide 1.000 0.120 Pass/<LOD
DDVP (Dichlorvos) 1.000 0.240 Pass/<LOD Diazinon 0.200 0.120 Pass/<LOD
Dimethoate 0.200 0.120 Pass/<LOD Ethoprophos 0.200 0.120 Pass/<LOD
Etofenprox 0.400 0.120 Pass/<LOD Etoxazole 0.200 0.120 Pass/<LOD
Fenoxycarb 0.200 0.120 Pass/<LOD Fenpyroximate 0.400 0.120 Pass/<LOD
Fipronil 0.400 0.120 Pass/<LOD Flonicamid 1.000 0.120 Pass/<LOD
Fludioxonil 0.400 0.120 Pass/<LOD Hexythiazox 1.000 0.120 Pass/<LOD
Imazalil 0.200 0.120 Pass/<LOD Imidacloprid 0.400 0.120 Pass/<LOD
Kresoxim Methyl 0.400 0.120 Pass/<LOD Malathion 0.200 0.120 Pass/<LOD
Metalaxyl 0.200 0.120 Pass/<LOD Methiocarb 0.200 0.120 Pass/<LOD
Methomyl 0.400 0.120 Pass/<LOD Methyl Parathion 0.200 0.120 Pass/<LOD
MGK-264‡ 0.200 0.120 Pass/<LOD Myclobutanil 0.200 0.120 Pass/<LOD
Naled 0.500 0.120 Pass/<LOD Oxamyl 1.000 0.120 Pass/<LOD
Paclobutrazol 0.400 0.120 Pass/<LOD Permethrins† 0.200 0.120 Pass/<LOD
Phosmet 0.200 0.120 Pass/<LOD Piperonyl Butoxide 2.000 0.120 Pass/0.464
Prallethrin 0.200 0.120 Pass/<LOD Propiconazole 0.400 0.120 Pass/<LOD
Propoxur 0.200 0.120 Pass/<LOD Pyrethrins* 1.000 0.120 Pass/<LOD
Pyridaben 0.200 0.120 Pass/<LOD Spinosad***** 0.200 0.120 Pass/<LOD
Spiromesifen 0.200 0.120 Pass/<LOD Spirotetramat 0.200 0.120 Pass/<LOD
Spiroxamine‡ 0.400 0.120 Pass/<LOD Tebuconazole 0.400 0.120 Pass/<LOD
Thiacloprid 0.200 0.120 Pass/<LOD Thiamethoxam 0.200 0.120 Pass/<LOD
Trifloxystrobin 0.200 0.120 Pass/<LOD

* Pyrethrins are reported as the sum of Jasmolin I, Cinerin I, and Pyrethrin I
** Cyfluthrins are reported as the sum of isomers Cyfluthrin I, II, III, and IV
*** Cypermethrins are reported as the sum of isomers Cypermethrin I, II, III, and IV
**** Abamectin is reported as the sum of Avermectin B1a and Avermectin B1b
***** Spinosad is reported as the sum of Spinosyn A and Spinosyn D
† Permethrin and Prallethrin are reported as the sum of cis and trans isomers
‡ MGK-264 and Spiroximine are reported as the sum of isomers I and II
MRL - Maximum Residue Limit; LOD - Limit of Detection

This sample has not been tested according to OAR 333-007. These results should therefore be used for research and development or quality control purposes only.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 3

ORELAPIRON LaboratoriesNELAP Accredited Laboratory

ORELAP Cert No. 4092-002
OLCC No. 1002158CD2E

Pesticide QC Report

Testing Accreditation #: 4092-002 Test Certificate #: 114134-001
Client Name, Sample Details

Modern Moonshine
Junction city, OR 97448
Sample: Delta 9 Distillate
Type: Extract
Method: FE-52 (EN 15662 & AOAC 2007.01)

Test Conditions

Prepsheet ID#: ORPS190111a
Scale: Veritas-OR1
Temp: 22 °C
Baro PE: 995.8 hPa
Analyst: JDM
Technician: JDM

Sample ID#: 114134

Lot #: R&D
Batch #: 114134
Harvest/Process Date: 01/01/2019
Serving Size (g): 0.1
Package Weight (g):1
Date Received: 01/02/2019
Test Date: 01/12/2019

Sample #114134
Target Compound Name Method Blank (μg/g) QC Spike (μg/g) Matrix Spike (μg/g) Matrix Spike Duplicate (μg/g) MS recovery% MSD recovery% Relative Percent Difference (%) QC Flag
Acephate N.D. 1 1.17 1.22 117.00 122.00 4.18
Acequinocyl N.D. 1 0.554 0.589 55.40 58.90 6.12 LR
Acetamiprid N.D. 1 1.14 1.18 114.00 118.00 3.45
Aldicarb N.D. 1 1.06 1.08 106.00 108.00 1.87
Avermectin B1a N.D. 0.97 1.21 1.27 124.74 130.93 4.84
Azoxystrobin N.D. 1 1.19 1.24 119.00 124.00 4.12
Bifenazate N.D. 1 1.33 1.25 133.00 125.00 6.20 HR
Bifenthrin N.D. 1 0.937 0.942 93.70 94.20 0.53
Boscalid N.D. 1 1.19 1.22 119.00 122.00 2.49
Carbaryl N.D. 1 1.12 1.14 112.00 114.00 1.77
Carbofuran N.D. 1 1.14 1.16 114.00 116.00 1.74
Chlorantraniliprole N.D. 1 1.53 1.48 153.00 148.00 3.32 HR
Chlorfenapyr N.D. 1 0.945 0.95 94.50 95.00 0.53
Chlorpyriphos N.D. 1 0.999 1.07 99.90 107.00 6.86
Clofentezine N.D. 1 1.15 1.19 115.00 119.00 3.42
Cyfluthrin N.D. 1 1.12 1.12 112.00 112.00 0.00
Cypermethrin N.D. 1 1.12 1.12 112.00 112.00 0.00
Daminoside N.D. 1 0.891 0.911 89.10 91.10 2.22
Diazanon N.D. 1 1.27 1.26 127.00 126.00 0.79
Dichlorvos N.D. 1 1.07 1.11 107.00 111.00 3.67
Dimethoate N.D. 1 1.14 1.18 114.00 118.00 3.45
Ethoprophos N.D. 1 1.1 1.16 110.00 116.00 5.31
Etofenprox N.D. 1 0.858 0.903 85.80 90.30 5.11
Etoxazole N.D. 1 1.16 1.21 116.00 121.00 4.22
Fenoxycarb N.D. 1 1.47 1.4 147.00 140.00 4.88
Fenpyroximate N.D. 1 1.11 1.15 111.00 115.00 3.54
Fipronil N.D. 1 0.981 1.01 98.10 101.00 2.91
Flonicamid N.D. 1 1.13 1.17 113.00 117.00 3.48
Fludioxonil N.D. 1 0.93 0.967 93.00 96.70 3.90
Hexythiazox N.D. 1 0.85 0.904 85.00 90.40 6.16
Imazalil N.D. 1 0.938 1.6 93.80 160.00 52.17 HRQ
Imidacloprid N.D. 1 1.25 1.27 125.00 127.00 1.59
Kresoxim-methyl N.D. 1 1.56 1.52 156.00 152.00 2.60 HR
Malathion N.D. 1 1.33 1.38 133.00 138.00 3.69 HR
Metalaxyl N.D. 1 1.44 1.4 144.00 140.00 2.82 HR
Methiocarb N.D. 1 1.09 1.13 109.00 113.00 3.60
Methomyl N.D. 1 1.27 1.28 127.00 128.00 0.78
MGK-264 N.D. 1 1.1 1.07 110.00 107.00 2.76
Myclobutanil N.D. 1 1.27 1.42 127.00 142.00 11.15 HR
Naled (dibrom) N.D. 1 0.869 0.906 86.90 90.60 4.17
Oxamyl N.D. 1 1.17 1.19 117.00 119.00 1.69
Paclobutrazol N.D. 1 1.53 1.49 153.00 149.00 2.65 HR
Parathion-methyl N.D. 1 1.07 1.06 107.00 106.00 0.94
Permethrins N.D. 1 0.98 0.978 98.00 97.80 0.20
Phosmet N.D. 1 1.43 1.37 143.00 137.00 4.29 HR
Piperonyl butoxide N.D. 1 1.13 1.12 113.00 112.00 0.89
Prallethrin N.D. 1 1.34 1.32 134.00 132.00 1.50 HR
Propiconazole N.D. 1 1.29 1.25 129.00 125.00 3.15
Propoxur N.D. 1 1.12 1.17 112.00 117.00 4.37
Pyrethrin N.D. 0.65 0.94 0.798 144.62 122.77 16.34 HR
Pyridaben N.D. 1 0.955 0.978 95.50 97.80 2.38
SpinosynA N.D. 0.84 1.28 1.2 152.38 142.86 6.45
SpinosynD N.D. 0.16 0.244 0.264 152.50 165.00 7.87 HR
Spiromesifen N.D. 1 0.465 0.486 46.50 48.60 4.42 LR
Spirotetramat N.D. 1 1.32 1.25 132.00 125.00 5.45
Spiroxamine N.D. 1 1.41 1.42 141.00 142.00 0.71 HR
Tebuconazole N.D. 1 1.48 1.51 148.00 151.00 2.01 HR
Thiacloprid N.D. 1 1.25 1.3 125.00 130.00 3.92
Thiamethoxam N.D. 1 1.35 1.33 135.00 133.00 1.49 HR
Trifloxystrobin N.D. 1 1.17 1.18 117.00 118.00 0.85

N.D. = Not Detected

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 30%

R = indicates compound recovery of matrix spike was outside the methods acceptable limits. (60-120%) Low recovery could indicate there is actually more compound present than detected; while high recoveries should be scrutinized for possible fails as more compound may be detected than is actually residual on the sample.

This sample has not been tested according to OAR 333-007. These results should therefore be used for research and development or quality control purposes only.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 3 of 3