ORELAPIRON LaboratoriesNELAP Accredited Laboratory

ORELAP Cert No. 4092-002
OLCC No. 1002158CD2E

Marijuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 4092-002 Test Certificate #: 104138-001
Client Name, Sample Details

Halfway 2 Happy farms
Halfway, OR 97834
Sample: Flower Salve
Type: Infused Product
Method: FE04E-OR4 HPLC1100-OR1

Test Conditions

Prepsheet ID#: P180313
Scale: XS205-OR1
Temp: 19.8 °C
Baro Pressure: 995.4 hPa
Analyst: HRM
Technician: HRM

Sample ID#: 104138

Harvest/Process Date: 03/09/2018
Serving Size (g): 9.627
Package Weight (g):9.627
Date Received: 03/09/2018
Test Date: 03/13/2018

Sample #104138
Sample #104138 Sample #104138
Test
Compounds
CBDV CBDA CBD CBC CBG CBN THC THCA THCV   AVG CBD Decarb. % THC Decarb. %
Amount 0.05 mg/g 0.12 mg/g 0.01 mg/g 0.61 mg/g 0.43 mg/g 0.30 mg/g 17.02 mg/g 6.89 mg/g 0.08 mg/g THC 23.06 mg/g 8 71
LOQ 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g 0.0006 mg/g CBD 0.12 mg/g - -
Amount per Serving~ 0.48 mg/srv 1.16 mg/srv 0.10 mg/srv 5.87 mg/srv 4.14 mg/srv 2.89 mg/srv 163.85 mg/srv 66.33 mg/srv 0.77 mg/srv - Serving Size~ (g): 9.63 g -
Uncertainty 0.81% RPD 0.58% RPD 0.00% RPD 2.82% RPD 0.20% RPD 0.38% RPD ±5% RPD 0.54% RPD 0.38% RPD - - - -

Serving Size = 1 tin of Flower Salve = 9.63 Grams

LOQ = Limit of Quantitation

RPD is relative percent difference between the LCS and LCS duplicate.

%RPD = Relative Percent Difference; Min. Value = Minimum Detectable Amount, CFU = Colony Forming Units, N/D = Not Detected

*** Designated tests that use the method FE-45.

Total CBD or THC is the calculated sum of CBD or THC and the amount of CBD or THC derived from CBDA or THCA, respectively. These values are calculated by adding a molar correction factor of 0.877 to the CBDA or THCA value.

THC and CBD Decarb. % referes to the percentage of THC or CBD relative to THCA or CBDA, respectively

Avg. is the calculated average of total CBD or total THC for all increments tested.

This sample has not been tested according to OAR 333-007. These results should therefore be used for research and development or quality control purposes only.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 1 of 1

ORELAPIRON LaboratoriesNELAP Accredited Laboratory

ORELAP Cert No. 4092-002
OLCC No. 1002158CD2E

QC Report

Testing Accreditation #: 4092-002 Test Certificate #: 104138-001
Client Name, Sample Details

Halfway 2 Happy farms
Halfway, OR 97834
Sample: Flower Salve
Type: Infused Product
Method: FE04E-OR4 HPLC1100-OR1

Test Conditions

Prepsheet ID#: P180313
Scale: XS205-OR1
Temp: 19.8 °C
Baro Pressure: 995.4 hPa
Analyst: HRM
Technician: HRM

Sample ID#: 104138

Harvest/Process Date: 03/09/2018
Serving Size (g): 9.627
Package Weight (g):9.627
Date Received: 03/09/2018
Test Date: 03/13/2018

Sample #104138
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Cannabidivarin (CBDV) N.D. 4.99 4.93 98.80 4.97 99.60 0.81%
Cannabidiolic Acid (CBDA) N.D. 5.13 5.17 100.78 5.14 100.19 0.58%
Cannabigerol (CBG) N.D. 4.99 5.08 101.80 5.07 101.60 0.20%
Cannabidiol (CBD) N.D. 5.12 5.25 102.54 5.25 102.54 0.00%
Δ9-Tetrahydrocannabivarin (THCV) N.D. 5.19 5.34 102.89 5.32 102.50 0.38%
Cannabinol (CBN) N.D. 5.13 5.3 103.31 5.32 103.70 0.38%
Δ9-Tetrahydrocannabinol (THC) N.D. 6.3 7.17 113.81 7.39 117.30 3.02%
Cannabichromene (CBC) N.D. 6.03 6.64 110.12 6.83 113.27 2.82%
Tetrahydrocannabinolic acid (THCA) N.D. 5.22 5.59 107.09 5.56 106.51 0.54%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

This sample has not been tested according to OAR 333-007. These results should therefore be used for research and development or quality control purposes only.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

Joseph Rutowski Iron laboratories Joseph Rutowski

Iron Labs Oregon complies with 2009 TNI Environmental Laboratory Standards.

Tested by Iron Laboratories Oregon, 71 Centennial Loop Suite D Eugene, OR 97401

Page 2 of 1