IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

Marihuana Potency Analysis by
High Performance Liquid Chromatography

Testing Accreditation #: 77802 Test Certificate #: 139531-001
Client Name, Sample Details

Bloomfield Development Group ( MED:GR-C-000683)
Warren, MI 48089
Sample: Detroit Kush Cake
License: GR-C-000683
Type: Flower
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A405010001A773000001756
Metrc Source Pkg#: 1A4050100013CCD000000169
***Water Activity: 0.55

Test Conditions

Prepsheet ID#: MIP230127A
Scale: XS205-MI2
Temp: 23.8 °C
Baro Pressure: 961.2 hPa
Analyst: KEB
Technician: SKM

Sample ID#: 139531

Harvest/Process Date: 01/27/2023
Serving Size (g): 1
Date Received: 01/27/2023
Test Date: 01/27/2023
Valid Through: 01/28/2024
Report Issued: 01/31/2023

Sample #139531
Sample #139531 Sample #139531
Test
Compounds
Δ9-THC THCA Δ8-THC CBD CBDA CBG CBGA CBN CBC CBL THCV CBDV Total
Cannabinoids*
Total THC Total CBD Calc Max Total
Cannabinoids*
Amount (%) 3.4 21.2 N/D N/D N/D 0.2 0.7 N/D 0.1 N/D N/D N/D 25.5 22.0 0.0 22.8
Amount (mg/g) 34.3 211.7 N/D N/D N/D 1.7 6.57 N/D 0.9 N/D N/D N/D 255.17 220.0 0.0 228.3
Amount per Serving (mg) 34.3 211.7 N/D N/D N/D 1.7 6.57 N/D 0.9 N/D N/D N/D 255.2 Serving Size~ (g): 1.0
LOQ (mg/g) 0.79 0.79 0.79 0.79 0.79 0.79 0.79 0.79 0.79 0.79 0.79 0.79   %Decarb. THC CBD
±%RPD 4.70% 14.58% 0.93% 2.91% 0.94% 1.94% 0.66% 1.13% 1.26% 0.13% 1.21% 0.83% 14 N/A

Sample passes for water activity (sample fails if Aw>0.65).

LOQ = Limit of Quantitation; %RPD = Relative Percent Deviation; %RSD = Relative Standard Deviation; N/D = Not Detected

*Designates values that are not currently included in the accredited scope of Iron Laboratories.

*** Designates tests that use the method FE-45. FE-45 is performed using AOAC 966.02 and 32.004-32.009.

FE-45 has relative expanded (k=2) uncertainties of 1.098% for moisture, 1.754% for water activity for unprocessed plant materials, and 13.102% for water activity for infused products.

Vitamin E acetate analysis has a relative expanded (k=2) uncertainty of 18.614%.

Total THC and CBD is the calculated sum of Δ9-THC and Δ8-THC or CBD plus the amount of THC or CBD derived from THCA or CBDA, respectively.

These values are calculated by applying a molar correction factor of 0.877 to the THCA or the CBDA value.

Calc Max Total Cannabinoids is the sum of Total THC, Total CBD, CBN, Total CBG, CBC, THCV, CBDV, and CBL.

Total CBG is calculated as CBG plus CBGA*0.878 (the molar correction factor for CBGA to CBG conversion).

%Decarb. THC and CBD refer to the percentage of Δ9-THC or CBD relative to THCA or CBDA, respectively.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 1 of 2

IRON Laboratories     PJLA

ISO/IEC 17025:2017 Accredited

QC Report

Testing Accreditation #: 77802 Test Certificate #: 139531-001
Client Name, Sample Details

Bloomfield Development Group ( MED:GR-C-000683)
Warren, MI 48089
Sample: Detroit Kush Cake
License: GR-C-000683
Type: Flower
Method: FE04U12 HPLC-UV
Metrc Test Pkg#: 1A405010001A773000001756
Metrc Source Pkg#: 1A4050100013CCD000000169
***Water Activity: 0.55

Test Conditions

Prepsheet ID#: MIP230127A
Scale: XS205-MI2
Temp: 23.8 °C
Baro Pressure: 961.2 hPa
Analyst: KEB
Technician: SKM

Sample ID#: 139531

Harvest/Process Date: 01/27/2023
Serving Size (g): 1
Date Received: 01/27/2023
Test Date: 01/27/2023
Valid Through: 01/28/2024
Report Issued: 01/31/2023

Sample #139531
Target Compound Name Method Blank (μg/g) LCS Spike (μg/g) LCS Amount (μg/g) Percent Recovery (%) LCS LCS Duplicate Amount (μg/g) Percent Recovery (%) LCSD Relative Percent Difference (%) QC Flag
Δ9-THC 0 5.71054 5.55935 97.35 5.82709 102.04 4.70%
THCA 0 4.97169 4.95326 99.63 5.73198 115.29 14.58% HR
Δ8-THC 0 5.44649 5.37621 98.71 5.42622 99.63 0.93%
CBD 0 5.22319 5.21131 99.77 5.06195 96.91 2.91%
CBDA 0 4.90076 4.88561 99.69 4.84012 98.76 0.94%
CBG 0 5.67488 5.60377 98.75 5.49593 96.85 1.94%
CBGA 0 5.4018 5.37451 99.49 5.33932 98.84 0.66%
CBN 0 5.04432 5.04805 100.07 4.99139 98.95 1.13%
CBC 0 5.34515 5.33143 99.74 5.2647 98.49 1.26%
CBCA N/A 0 0 0.00 0 0.00 0%
CBL 0 5.42353 5.25606 96.91 5.26274 97.04 0.13%
THCV 0 5.35708 5.37079 100.26 5.30602 99.05 1.21%
CBDV 0 5.53707 5.51693 99.64 5.4715 98.82 0.83%

N.D. = Not Detected

LR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) Low recovery should be scrutinized for possible fail as it could indicate more compound present than is detected.

I = indicates that an amount of an interfering compound greater than the methods limit of detection was detected in the method blank sample. May indicate contamination of analytical system or consumables.

Q = indicates that the relative percent diference of two identicly prepared Matrix Spike samples for a target analyte was greater than 20%

HR = indicates compound recovery of matrix spike was outside the methods acceptable limits. (70-130%) high recoveries should be scrutinized for passing as more compound may be detected than is actually present in the sample.

Sample was sampled and tested in accordance with the Safety Compliance Facility Sampling and Testing Information.

This certificate shall not be reproduced except in full, without written approval of Iron Laboratories, LLC.

KEB - LM Iron laboratories KEB - LM

Iron Laboratories, LLC is an ISO/IEC 17025:2017 Testing Laboratory laboratory accredited by (PJLA) Perry Johnson Laboratory Accreditation, Certificate No. 77802

Tested by Iron Laboratories Michigan, 1825 E. West Maple Walled Lake, MI 48390

Page 2 of 2